FDA Adverse Event Injury Summary report: N

PROTECTA XT CRT-D

MDR report key: 3860263 · Received June 10, 2014

Report

Report Number
3004209178-2014-10598
Event Type
Injury
Date Received
June 10, 2014
Date of Event
March 21, 2014
Report Date
March 21, 2014
Manufacturer
MEDTRONIC MED REL MEDTRONIC PUERTO RICO
Product Code
NIK
PMA / PMN Number
P010031
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PRODUCT EVENT SUMMARY: THE DEVICE WAS RETURNED AND ANALYZED. THE DEVICE MET 84% OF EXPECTED LONGEVITY. WITHOUT THE HISTORY OF THE PROGRAMMED SETTINGS THROUGHOUT ITS SERVICE LIFE, THERE IS NO WAY TO DETERMINE WHY THE LONGEVITY DID NOT MATCH THE PREDICTED MODEL. CONCOMITANT PRODUCTS: 6943-65, LEAD, IMPLANTED: (B)(6) 2001; 419388, LEAD, IMPLANTED: (B)(6) 2009. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PHYSICIAN WAS CONCERNED THE DEVICE BATTERY DEPLETED EARLY. THE DEVICE WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
342936 PROTECTA XT CRT-D DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CARDIAC RESYNCHRONIZATIO NIK MEDTRONIC MED REL MEDTRONIC PUERTO RICO D314TRG

Patients

Seq Age Sex Outcome Treatment
1 00084 YR Hospitalization| R 5076-52, LEAD