MDR report key: 3860180 · Received June 10, 2014

Report

Report Number
3008973940-2014-00163
Event Type
Malfunction
Date Received
June 10, 2014
Date of Event
March 10, 2014
Report Date
March 10, 2014
Manufacturer
MEDTRONIC SINGAPORE OPERATIONS
Product Code
DTB
PMA / PMN Number
P930039
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

A GOOD FAITH EFFORT WILL BE MADE TO OBTAIN THE APPLICABLE INFORMATION RELEVANT TO THE REPORT. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Additional Manufacturer Narrative · 1

THIS EVENT OCCURRED OUTSIDE THE US WHERE THE SAME MODEL IS DISTRIBUTED. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. PRODUCT EVENT SUMMARY: THE FULL LEAD WAS RETURNED, ANALYZED AND THE DISTAL CONDUCTOR OF THE LEAD WAS EXTRINSICALLY OVER-ROTATED. THE DISTAL LV (LOW VOLTAGE) ELECTRODE OF THE LEAD WAS COVERED IN BODY TISSUE/FIBROTIC GROWTH. VISUAL SUMMARY ANALYSIS OF THE LEAD INDICATED DAMAGE AT IMPLANT AND STRETCHING. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING THE IMPLANT PROCEDURE, THE SCREW WAS INSIDE THE HOUSING BEFORE IMPLANT. AFTER VARIOUS REPOSITIONING TRIES IT WAS DISCOVERED THE SCREW WAS OUTSIDE THE HOUSING. THE LEAD WAS ATTEMPTED AND NOT USED. ANOTHER LEAD WAS USED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
337492 CAPSUREFIX NOVUS ELECTRODE, PACEMAKER, PERMANENT DTB MEDTRONIC SINGAPORE OPERATIONS 5076-58

Patients

Seq Age Sex Outcome Treatment
1 00043 YR