CONSULTA CRT-D
Report
- Report Number
- 3004209178-2014-10604
- Event Type
- Injury
- Date Received
- June 10, 2014
- Date of Event
- March 12, 2014
- Report Date
- March 12, 2014
- Manufacturer
- MEDTRONIC MED REL MEDTRONIC PUERTO RICO
- Product Code
- NIK
- PMA / PMN Number
- P010031
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- RI, US
- Reporter Occupation
- PHYSICIAN
Narratives
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PRODUCT EVENT SUMMARY: THE DEVICE WAS RETURNED AND ANALYZED. THE DEVICE MET 98% OF EXPECTED LONGEVITY. WITHOUT THE HISTORY OF THE PROGRAMMED SETTINGS THROUGHOUT ITS SERVICE LIFE, THERE IS NO WAY TO DETERMINE WHY THE LONGEVITY DID NOT MATCH THE PREDICTED MODEL. CONCOMITANT PRODUCTS: MODEL: 4470, COMPETITOR IMPLANTABLE PACING LEAD; IMPLANT: (B)(6) 2006; MODEL: 419478, IMPLANTABLE PACING LEAD; IMPLANT: (B)(6) 2006.
IT WAS REPORTED THAT EARLY BATTERY DEPLETION OF A BI-V IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) RESULTED IN AN INTERVENTION TO EXPLANT AND REPLACE THE DEVICE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 336941 | CONSULTA CRT-D | DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CARDIAC RESYNCHRONIZATIO | NIK | MEDTRONIC MED REL MEDTRONIC PUERTO RICO | D224TRK |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 00068 YR | Hospitalization| R | 694965, IMPLANTABLE TACHY LEAD |