CONCERTO II CRT-D
Report
- Report Number
- 3004209178-2014-10616
- Event Type
- Injury
- Date Received
- June 10, 2014
- Date of Event
- March 25, 2014
- Report Date
- March 25, 2014
- Manufacturer
- MEDTRONIC MED REL MEDTRONIC PUERTO RICO
- Product Code
- NIK
- PMA / PMN Number
- P010031
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MO, US
- Reporter Occupation
- PHYSICIAN
Narratives
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PRODUCT EVENT SUMMARY: THE DEVICE WAS RETURNED AND ANALYZED. ANALYSIS REVEALED THE RETURNED DEVICE DID NOT DETECT ANY PERFORMANCE ISSUES, BUT THE CALCULATED LONGEVITY RESULT OF 91% OF EXPECTED WAS LESS THAN THE PREDICTED VALUE BASED ON THE AVAILABLE PROGRAMMED PARAMETERS. IN THE ABSENCE OF THE COMPLETE PROGRAMMING HISTORY, IT IS NOT POSSIBLE TO DETERMINE WHY THE LONGEVITY DID NOT MATCH THE PREDICTED MODEL. CONCOMITANT PRODUCTS: 5076-52, LEAD, IMPLANTED: (B)(6) 2011; 419488, LEAD, IMPLANTED: (B)(6) 2011. (B)(4).
IT WAS REPORTED THAT THE DEVICE HAD UNEXPECTED LONGEVITY. THE DEVICE WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 337278 | CONCERTO II CRT-D | DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CARDIAC RESYNCHRONIZATIO | NIK | MEDTRONIC MED REL MEDTRONIC PUERTO RICO | D274TRK |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 00055 YR | Hospitalization| R | 694765, LEAD |