FDA Adverse Event Injury Summary report: N

SPRINT QUATTRO SECURE

MDR report key: 3860140 · Received June 10, 2014

Report

Report Number
2649622-2014-05551
Event Type
Injury
Date Received
June 10, 2014
Date of Event
January 12, 2014
Report Date
March 31, 2014
Manufacturer
MPRI
Product Code
LWS
PMA / PMN Number
P920015
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
ND, US
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. CONCOMITANT MEDICAL PRODUCTS: 5076 IMPLANTABLE PACING LEAD, (B)(6) 2008. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT THE RIGHT VENTRICULAR LEAD HAD T-WAVE OVERSENSING (TWOS) POST PACE. IT WAS NOTED THAT RATE ADAPTIVE AV IS ON AND THE PATIENT HAS A SINUS RATE OF 107 BPM. REPROGRAMMING WAS DISCUSSED TO CONSIDER TURNING OFF RA-AV IN THIS PATIENT WITH INTACT CONDUCTION. THE LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
337169 SPRINT QUATTRO SECURE DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER LWS MPRI 694765

Patients

Seq Age Sex Outcome Treatment
1 00071 YR Required Intervention 7278 ICD