FDA Adverse Event
Injury
Summary report: N
SPRINT QUATTRO SECURE
MDR report key: 3860140
·
Received June 10, 2014
Report
- Report Number
- 2649622-2014-05551
- Event Type
- Injury
- Date Received
- June 10, 2014
- Date of Event
- January 12, 2014
- Report Date
- March 31, 2014
- Manufacturer
- MPRI
- Product Code
- LWS
- PMA / PMN Number
- P920015
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- ND, US
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
Additional Manufacturer Narrative · 1
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. CONCOMITANT MEDICAL PRODUCTS: 5076 IMPLANTABLE PACING LEAD, (B)(6) 2008. (B)(4).
Description of Event or Problem · 1
IT WAS REPORTED THAT THE RIGHT VENTRICULAR LEAD HAD T-WAVE OVERSENSING (TWOS) POST PACE. IT WAS NOTED THAT RATE ADAPTIVE AV IS ON AND THE PATIENT HAS A SINUS RATE OF 107 BPM. REPROGRAMMING WAS DISCUSSED TO CONSIDER TURNING OFF RA-AV IN THIS PATIENT WITH INTACT CONDUCTION. THE LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 337169 | SPRINT QUATTRO SECURE | DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER | LWS | MPRI | 694765 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 00071 YR | Required Intervention | 7278 ICD |