FDA Adverse Event Injury Summary report: N

CAPSUREFIX

MDR report key: 3860131 · Received June 10, 2014

Report

Report Number
2649622-2014-05571
Event Type
Injury
Date Received
June 10, 2014
Date of Event
March 12, 2014
Report Date
March 12, 2014
Manufacturer
MPRI
Product Code
DTB
PMA / PMN Number
P930039
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IN, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. CONCOMITANT PRODUCT: 4076-58 LEAD, (B)(6) 2013. (B)(4). EVALUATION SUMMARY: ANALYSIS WAS PERFORMED AND NO ANOMALIES WERE FOUND, HOWEVER THE DISTAL LV (LOW VOLTAGE) ELECTRODE OF THE LEAD WAS COVERED IN BLOOD, THE DISTAL LV (LOW VOLTAGE) ELECTRODE OF THE LEAD WAS COVERED IN BODY TISSUE/FIBROTIC GROWTH, AND VISUAL SUMMARY ANALYSIS OF THE LEAD INDICATED APPARENT EXPLANT DAMAGE; FULL LEAD RETURNED AND ANALYZED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE RIGHT ATRIAL (RA) LEAD WAS FOUND TO HAVE NO CAPTURE AND OVERSENSING P-WAVES. THE LEAD WAS EXTRACTED AND A NEW LEAD WAS IMPLANTED SUCCESSFULLY. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
337075 CAPSUREFIX ELECTRODE, PACEMAKER, PERMANENT DTB MPRI 5568-45

Patients

Seq Age Sex Outcome Treatment
1 00086 YR Hospitalization| R ADDR01 IPG