CAPSUREFIX NOVUS
Report
- Report Number
- 2649622-2014-05582
- Event Type
- Injury
- Date Received
- June 10, 2014
- Date of Event
- March 14, 2014
- Report Date
- March 14, 2014
- Manufacturer
- MPRI
- Product Code
- DTB
- PMA / PMN Number
- P930039
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TN, US
- Reporter Occupation
- PHYSICIAN
Narratives
PRODUCT EVENT SUMMARY: THE PROXIMAL SEGMENT OF THE LEAD WAS RETURNED AND ANALYZED AND NO ANOMALIES WERE FOUND. THE OUTER INSULATION OF THE LEAD DEVELOPED A COSMETIC DEPRESSION WHILE IN VIVO. THERE WERE NO ANOMALIES OBSERVED REGARDING THE RETURNED LEAD PROXIMAL SEGMENT. ALL ELECTRICAL TESTING WAS WITHIN SPECIFIED PARAMETERS. THE FULL LEAD WAS NOT RETURNED. AN IN-VIVO INSULATION BREACH WAS NOT OBSERVED.
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. CONCOMITANT MEDICAL PRODUCTS: ADDR01 IPG IMPLANTED: (B)(6) 2009. (B)(4).
IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.
IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD HAD LOW IMPEDANCES DUE TO AN INSULATION BREACH. IT WAS FURTHER REPORTED THAT THE PATIENT EXPERIENCED POCKET STIMULATION. THE LEAD WAS EXPLANTED AND REPLACED. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 340532 | CAPSUREFIX NOVUS | ELECTRODE, PACEMAKER, PERMANENT | DTB | MPRI | 5076-58 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 00058 YR | Hospitalization| R | 5076-52 LEAD |