FDA Adverse Event Injury Summary report: N

CAPSUREFIX NOVUS

MDR report key: 3860088 · Received June 10, 2014

Report

Report Number
2649622-2014-05582
Event Type
Injury
Date Received
June 10, 2014
Date of Event
March 14, 2014
Report Date
March 14, 2014
Manufacturer
MPRI
Product Code
DTB
PMA / PMN Number
P930039
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TN, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

PRODUCT EVENT SUMMARY: THE PROXIMAL SEGMENT OF THE LEAD WAS RETURNED AND ANALYZED AND NO ANOMALIES WERE FOUND. THE OUTER INSULATION OF THE LEAD DEVELOPED A COSMETIC DEPRESSION WHILE IN VIVO. THERE WERE NO ANOMALIES OBSERVED REGARDING THE RETURNED LEAD PROXIMAL SEGMENT. ALL ELECTRICAL TESTING WAS WITHIN SPECIFIED PARAMETERS. THE FULL LEAD WAS NOT RETURNED. AN IN-VIVO INSULATION BREACH WAS NOT OBSERVED.

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. CONCOMITANT MEDICAL PRODUCTS: ADDR01 IPG IMPLANTED: (B)(6) 2009. (B)(4).

Additional Manufacturer Narrative · 1

IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD HAD LOW IMPEDANCES DUE TO AN INSULATION BREACH. IT WAS FURTHER REPORTED THAT THE PATIENT EXPERIENCED POCKET STIMULATION. THE LEAD WAS EXPLANTED AND REPLACED. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
340532 CAPSUREFIX NOVUS ELECTRODE, PACEMAKER, PERMANENT DTB MPRI 5076-58

Patients

Seq Age Sex Outcome Treatment
1 00058 YR Hospitalization| R 5076-52 LEAD