FDA Adverse Event Injury Summary report: N

VIVA QUAD XT

MDR report key: 3860071 · Received June 10, 2014

Report

Report Number
9614453-2014-01203
Event Type
Injury
Date Received
June 10, 2014
Date of Event
April 9, 2014
Report Date
April 10, 2014
Manufacturer
IPG MFG SWITZERLAND
Product Code
NIK
PMA / PMN Number
P010031
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THIS MODEL NUMBER IS NOT APPROVED FOR DISTRIBUTION IN THE U.S., HOWEVER, IT IS SIMILAR TO A DEVICE MARKETED IN THE U.S. THE EVENT OCCURRED OUTSIDE THE US AND IS BEING REPORTED DUE TO AN ALLEGED MALFUNCTION. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. PRODUCTS: 439888 LEAD IMPLANTED: 2014 (B)(6); 5554-53 LEAD IMPLANTED: 2014 (B)(6). (B)(4).

Additional Manufacturer Narrative · 1

PRODUCT EVENT SUMMARY: THE DEVICE WAS NOT RETURNED FOR ANALYSIS. HOWEVER, PERFORMANCE DATA COLLECTED FROM THE DEVICE WAS RECEIVED AND ANALYZED. ANALYSIS OF THE DEVICE MEMORY INDICATED A POWER ON RESET (POR) OCCURRED. ANALYST COMMENTED, POR OCCURRED ON (B)(6) 2014.

Additional Manufacturer Narrative · 1

IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) ENCOUNTERED A POWER ON RESET. THE PATIENT WAS HOSPITALIZED. THE DEVICE REMAINS IN USE, HOWEVER, THERE IS PLANNED INTERVENTION. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
339363 VIVA QUAD XT DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CARDIAC RESYNCHRONIZATIO NIK IPG MFG SWITZERLAND DTBA2QQ

Patients

Seq Age Sex Outcome Treatment
1 00067 YR Hospitalization| R 6947M62 LEAD