FDA Adverse Event
Injury
Summary report: N
CONCERTO II CRT-D
MDR report key: 3860022
·
Received June 10, 2014
Report
- Report Number
- 3004209178-2014-10650
- Event Type
- Injury
- Date Received
- June 10, 2014
- Date of Event
- March 6, 2014
- Report Date
- March 6, 2014
- Manufacturer
- MEDTRONIC MED REL MEDTRONIC PUERTO RICO
- Product Code
- NIK
- PMA / PMN Number
- P010031
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- KY, US
- Reporter Occupation
- PHYSICIAN
Narratives
Additional Manufacturer Narrative · 1
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PRODUCT EVENT SUMMARY : THE DEVICE WAS RETURNED AND ANALYZED. THE DEVICE MET 83% OF EXPECTED LONGEVITY. WITHOUT THE HISTORY OF THE PROGRAMMED SETTINGS THROUGHOUT ITS SERVICE LIFE, THERE IS NO WAY TO DETERMINE WHY THE LONGEVITY DID NOT MATCH THE PREDICTED MODEL. CONCOMITANT PRODUCTS: 5076-52 LEAD, IMPLANTED: (B)(6) 2010; 4194-78 LEAD, IMPLANTED: (B)(6) 2010. (B)(4).
Description of Event or Problem · 1
IT WAS REPORTED THAT THE DEVICE HAD EARLY BATTERY DEPLETION, WITH A BATTERY LIFE OF LESS THAN FOUR YEARS. THE DEVICE WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 340205 | CONCERTO II CRT-D | DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CARDIAC RESYNCHRONIZATIO | NIK | MEDTRONIC MED REL MEDTRONIC PUERTO RICO | D274TRK |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 00083 YR | Hospitalization| R | 6947-65 LEAD |