FDA Adverse Event Injury Summary report: N

CAPSUREFIX

MDR report key: 3860011 · Received June 10, 2014

Report

Report Number
2649622-2014-05638
Event Type
Injury
Date Received
June 10, 2014
Date of Event
April 15, 2014
Report Date
April 15, 2014
Manufacturer
MPRI
Product Code
DTB
PMA / PMN Number
P930039
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. CONCOMITANT MEDICAL PRODUCTS: A 694965 IMPLANTABLE TACHY LEAD, IMPLANTED: (B)(6) 2004. (B)(4).

Additional Manufacturer Narrative · 1

PRODUCT EVENT SUMMARY: THE LEAD WAS RETURNED AND ANALYSIS REVEALED THE INNER INSULATION OF THE LEAD DEVELOPED A BREACH DUE TO MIO (METAL ION OXIDATION) WHILE IN VIVO.

Additional Manufacturer Narrative · 1

IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE RIGHT ATRIAL (RA) LEAD HAD A POSSIBLE FRACTURE. THE LEAD WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
340737 CAPSUREFIX ELECTRODE, PACEMAKER, PERMANENT DTB MPRI 4568-53

Patients

Seq Age Sex Outcome Treatment
1 00063 YR Hospitalization| R D224DRG ICD