CAPSUREFIX
Report
- Report Number
- 2649622-2014-05627
- Event Type
- Malfunction
- Date Received
- June 10, 2014
- Date of Event
- March 27, 2014
- Report Date
- March 27, 2014
- Manufacturer
- MPRI
- Product Code
- DTB
- PMA / PMN Number
- P930039
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TU
- Reporter Occupation
- PHYSICIAN
Narratives
PRODUCT EVENT SUMMARY: THE FULL LEAD WAS RETURNED AND ANALYZED. THE DISTAL CONDUCTOR OF THE LEAD WAS EXTRINSICALLY OVER-ROTATED. VISUAL SUMMARY ANALYSIS OF THE LEAD INDICATED STRETCHING AND DAMAGE AT IMPLANT.
CORRECTED INFORMATION: NO EVAL EXPLAIN CODE. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.
THIS EVENT OCCURRED OUTSIDE THE US WHERE THE SAME MODEL IS DISTRIBUTED. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. (B)(4).
IT WAS REPORTED THAT DURING THE IMPLANT ATTEMPT, THE PHYSICIAN ATTEMPTED TO FIXATE THE ATRIAL LEAD, BUT WAS UNSUCCESSFUL. THE LEAD WAS REMOVED AND TESTED ON THE TABLE, AND A HELIX ISSUE WAS OBSERVED. THE LEAD WAS NOT USED AND ANOTHER LEAD WAS IMPLANTED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 340142 | CAPSUREFIX | ELECTRODE, PACEMAKER, PERMANENT | DTB | MPRI | 5568-53 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 00043 YR |