FDA Adverse Event Malfunction Summary report: N

CAPSUREFIX

MDR report key: 3860003 · Received June 10, 2014

Report

Report Number
2649622-2014-05627
Event Type
Malfunction
Date Received
June 10, 2014
Date of Event
March 27, 2014
Report Date
March 27, 2014
Manufacturer
MPRI
Product Code
DTB
PMA / PMN Number
P930039
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TU
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

PRODUCT EVENT SUMMARY: THE FULL LEAD WAS RETURNED AND ANALYZED. THE DISTAL CONDUCTOR OF THE LEAD WAS EXTRINSICALLY OVER-ROTATED. VISUAL SUMMARY ANALYSIS OF THE LEAD INDICATED STRETCHING AND DAMAGE AT IMPLANT.

Additional Manufacturer Narrative · 1

CORRECTED INFORMATION: NO EVAL EXPLAIN CODE. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Additional Manufacturer Narrative · 1

THIS EVENT OCCURRED OUTSIDE THE US WHERE THE SAME MODEL IS DISTRIBUTED. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING THE IMPLANT ATTEMPT, THE PHYSICIAN ATTEMPTED TO FIXATE THE ATRIAL LEAD, BUT WAS UNSUCCESSFUL. THE LEAD WAS REMOVED AND TESTED ON THE TABLE, AND A HELIX ISSUE WAS OBSERVED. THE LEAD WAS NOT USED AND ANOTHER LEAD WAS IMPLANTED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
340142 CAPSUREFIX ELECTRODE, PACEMAKER, PERMANENT DTB MPRI 5568-53

Patients

Seq Age Sex Outcome Treatment
1 00043 YR