FDA Adverse Event Death Summary report: N

8400STI

MDR report key: 38590 · Received September 13, 1996

Report

Report Number
2021710-1996-00001
Event Type
Death
Date Received
September 13, 1996
Date of Event
August 12, 1996
Report Date
August 13, 1996
Manufacturer
BIRD PRODUCTS CORP.
Product Code
CBK
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
RESPIRATORY THERAPIST

Narratives

Additional Manufacturer Narrative · 1

THE REPORTED EVENT WAS TRACED TO THE PT CIRCUIT SET-UP BEING USED. THIS DEVICE DID NOT MALFUNCTION AND OPERATED PER SPECIFICATIONS WHEN CHECKED. THE PT CIRCUIT SET-UP USED FOR THIS PT INCLUDED A NUMBER OF FILTERS IN-LINE IN THE INSPIRATORY LEG OF THIS CIRCUIT. THESE FILTERS CAUSED ADDITIONAL RESISTANCE IN THE PT CIRCUIT. THE PT HAD TO OVERCOME THIS RESISTANCE IN ORDER TO DRAW AIR FROM THE VENTILATOR. WHEN THESE FILTERS WERE REMOVED FROM THE PT CIRCUIT, THE VENTILATOR OPERATED NORMALLY. THIS HOSPITAL WAS INSERVICED IN THE PROPER TECHNIQUE FOR PRESSURE AND FLOW TRIGGERING BY CO'S CLINICAL STAFF. SECTION F WAS ENTERED BY THE MFR BASED FON INFROMATION PROVIDED BY THE HOSP.

Description of Event or Problem · 1

UNIT WAS NOT ADEQUATELY SENSING THE INSPIRATORY EFFORT OF THIS PT. THIS PT'S WORK OF BREATHING WAS REPORTED TO BE MARKEDLY HIGH. THIS HOSP REPORTED A PROGRESSIVE DETERIORATION OF THIS PT. THIS PT CODED AND EXPIRED EARLY THE NEXT MORNING. THIS HOSP IS NOT SURE IF THE DEATH OF THIS PT IS RELATED TO THE INCREASE WORK OF BREATHING.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 8400STI VOLUME VENTILATOR CBK BIRD PRODUCTS CORP. 15404 NA

Patients

Seq Age Sex Outcome Treatment
1 70 YR Death HUDSON RCI KRATON VENTILATOR TUBING SET(690-40)| GIBECK BACTERIAL FILTER (#1910)| GIBECK HUMID-VENT FILTER RT30+ (#19992)