FDA Adverse Event Malfunction Summary report: N

EASYSTAND STRAPSTAND

MDR report key: 3856174 · Received April 25, 2014

Report

Report Number
2183634-2014-00002
Event Type
Malfunction
Date Received
April 25, 2014
Date of Event
March 26, 2014
Report Date
March 26, 2014
Manufacturer
ALTIMATE MEDICAL, INC.
Product Code
ION
PMA / PMN Number
K885343
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MO, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

THE ADJUSTABLE SEAT STRAP WAS RETURNED TO ALTIMATE AND REVIEWED. UPON REVIEW, IT WAS NOTED THAT THE INSERTED STRAP IN THE SEAT MISSED THE LAST ROW OF STITCHING. THIS APPEARS TO HAVE BEEN CAUSED BY INCONSISTENT MANUFACTURING PROCESSES AT THE SUPPLIER WHO MANUFACTURES THESE STRAPS. ALTIMATE MEDICAL HAS UPDATED THE PRODUCT SPECIFICATIONS TO ADD ADD'L DETAILED INFO TO THE PRODUCT SPECIFICATIONS FOR THIS COMPONENT AND HAS COMMUNICATED WITH THE SUPPLIER REGARDING THEIR QUALITY CONTROL PROCESS. ALTIMATE HAS ALSO IMPLEMENTED A RECEIVING INSPECTION PROCESS ON THIS COMPONENT IN ADDITION TO THE CURRENT INSPECTION PROCESS PERFORMED DURING THE ASSEMBLY AND SHIPPING PROCESS.

Description of Event or Problem · 1

RECEIVED A CALL FROM A WOMAN WHO PURCHASED A STRAPSTAND IN (B)(6) 2012 FOR HER HUSBAND. SHE SAID HER AND A HELPER WERE PUTTING HIM IN THE UNIT WHEN ONE SIDE OF THE STRAP TORE AWAY FROM THE PADDED SEAT BODY OF THE XL ADJUSTABLE SEAT STRAP ON AN EASYSTAND STRAPSTAND. THE USER WAS NOT INJURED BUT STARTLED WHEN IT OCCURRED. BOTH OF THEM WERE ON EACH SIDE OF HIM AND HE FELL BACK INTO HIS WHEELCHAIR, WHICH IS POSITIONED UP TO THE STANDING FRAME, SO IT'S AVAILABLE WHEN THE USER GOES BACK TO THE SEATED POSITION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
252306 EASYSTAND STRAPSTAND 890.5370 ION ALTIMATE MEDICAL, INC. STRAPSTAND NA

Patients

Seq Age Sex Outcome Treatment
1