EASYSTAND STRAPSTAND
Report
- Report Number
- 2183634-2014-00002
- Event Type
- Malfunction
- Date Received
- April 25, 2014
- Date of Event
- March 26, 2014
- Report Date
- March 26, 2014
- Manufacturer
- ALTIMATE MEDICAL, INC.
- Product Code
- ION
- PMA / PMN Number
- K885343
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MO, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
THE ADJUSTABLE SEAT STRAP WAS RETURNED TO ALTIMATE AND REVIEWED. UPON REVIEW, IT WAS NOTED THAT THE INSERTED STRAP IN THE SEAT MISSED THE LAST ROW OF STITCHING. THIS APPEARS TO HAVE BEEN CAUSED BY INCONSISTENT MANUFACTURING PROCESSES AT THE SUPPLIER WHO MANUFACTURES THESE STRAPS. ALTIMATE MEDICAL HAS UPDATED THE PRODUCT SPECIFICATIONS TO ADD ADD'L DETAILED INFO TO THE PRODUCT SPECIFICATIONS FOR THIS COMPONENT AND HAS COMMUNICATED WITH THE SUPPLIER REGARDING THEIR QUALITY CONTROL PROCESS. ALTIMATE HAS ALSO IMPLEMENTED A RECEIVING INSPECTION PROCESS ON THIS COMPONENT IN ADDITION TO THE CURRENT INSPECTION PROCESS PERFORMED DURING THE ASSEMBLY AND SHIPPING PROCESS.
RECEIVED A CALL FROM A WOMAN WHO PURCHASED A STRAPSTAND IN (B)(6) 2012 FOR HER HUSBAND. SHE SAID HER AND A HELPER WERE PUTTING HIM IN THE UNIT WHEN ONE SIDE OF THE STRAP TORE AWAY FROM THE PADDED SEAT BODY OF THE XL ADJUSTABLE SEAT STRAP ON AN EASYSTAND STRAPSTAND. THE USER WAS NOT INJURED BUT STARTLED WHEN IT OCCURRED. BOTH OF THEM WERE ON EACH SIDE OF HIM AND HE FELL BACK INTO HIS WHEELCHAIR, WHICH IS POSITIONED UP TO THE STANDING FRAME, SO IT'S AVAILABLE WHEN THE USER GOES BACK TO THE SEATED POSITION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 252306 | EASYSTAND STRAPSTAND | 890.5370 | ION | ALTIMATE MEDICAL, INC. | STRAPSTAND | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|