ADVIA CENTAUR XP RUBELLA G (RUB G) ASSAY
Report
- Report Number
- 1219913-2014-00140
- Event Type
- Malfunction
- Date Received
- June 9, 2014
- Date of Event
- May 15, 2014
- Report Date
- May 15, 2014
- Manufacturer
- SIEMENS HEALTHCARE DIAGNOSTICS, INC.
- Product Code
- LFX
- PMA / PMN Number
- K003412
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IA, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
A SIEMENS FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED TO THE CUSTOMER SITE ON (B)(4) 2014. THE FSE REPLACED THE PINCH VALVE TUBING AND WASH BLOCKS. THE FSE CHECKED THE PROBE DISPENSE AND REPLACED THE MANIFOLD ASSEMBLY. THE CUSTOMER VERIFIED THE QUALITY CONTROL. ON THE FOLLOW-UP VISIT ON (B)(4) 2014, THE FSE REPLACED THE ASPIRATE MANIFOLD ASSEMBLY, THE ACID AND BASE PUMP, AND BLEACH VALVES. THE FSE PRIMED THE ACID AND BASE AND CHECKED THE SYSTEM FOR BLEACH. THE CUSTOMER RECALIBRATED ASSAYS. SOME OF THE ASSAY CALIBRATIONS FAILED. THE ACID AND BASE WERE REPLACED. THE RUBELLA G NEGATIVE QC WAS FINE POST-SERVICE. HOWEVER, ON (B)(4) 2014, THE NEGATIVE QC WAS RESULT WAS 4.2 (RANGE 0 TO 3). THE REAGENT READYPACK WAS NOTED TO HAVE MATERIAL ON THE BOTTOM AFTER MIXING. FRESH QC MATERIAL WAS TESTED ON THE SAME REAGENT READYPACK AND THE RESULT WAS 4.5. PATIENT SAMPLES WERE NOT RUN WHEN THE NEGATIVE QC WAS OUT OF RANGE. A NEW REAGENT READYPACK WAS MIXED AND THE QC WAS RUN. THE NEGATIVE QC RESULT WAS 0.2 (WITHIN RANGE). THE CAUSE FOR QC OUT OF RANGE IS POSSIBLE SUSPECT REAGENT READYPACK. THE CAUSE FOR THE DISCORDANT RUBELLA G PATIENT SAMPLE RESULT IS UNKNOWN. THE INSTRUMENT IS PERFORMING WITHIN SPECIFICATION. NO FURTHER EVALUATION OF THE DEVICE IS REQUIRED. THE IFU STATES IN THE QUALITY CONTROL SECTION: TAKING CORRECTIVE ACTION IF THE QUALITY CONTROL RESULTS DO NOT FALL WITHIN THE EXPECTED VALUES OR WITHIN THE LABORATORY'S ESTABLISHED VALUES, DO NOT REPORT RESULTS. TAKE THE FOLLOWING ACTIONS: -VERIFY THAT THE MATERIALS ARE NOT EXPIRED. -VERIFY THAT REQUIRED MAINTENANCE WAS PERFORMED. -VERIFY THAT THE ASSAY WAS PERFORMED ACCORDING TO THE INSTRUCTIONS FOR USE.
A FALSE POSITIVE ADVIA CENTAUR XP RUBELLA G (RUB G) RESULT WAS OBTAINED FOR A PATIENT SAMPLE. THE PREVIOUSLY NEGATIVE PATIENT SAMPLE WAS TESTED AS PART OF TROUBLESHOOTING THE NEGATIVE QUALITY CONTROL (QC) OUT OF RANGE. THE REAGENT READYPACK WAS CALIBRATED AND THE QC WAS REPEATED. THE NEGATIVE QC WAS IN RANGE. ROUTINE PATIENT SAMPLES WERE NOT RUN WHILE THE QC WAS OUT OF RANGE. PATIENT TREATMENT WAS NOT PRESCRIBED OR ALTERED. THERE WAS NO REPORT OF ADVERSE HEALTH CONSEQUENCES DUE TO THE DISCORDANT RUBELLA G RESULT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 336127 | ADVIA CENTAUR XP RUBELLA G (RUB G) ASSAY | RUBELLA G IMMUNOASSAY | LFX | SIEMENS HEALTHCARE DIAGNOSTICS, INC. | N/A | 179 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |