FDA Adverse Event Malfunction Summary report: N

HOMECHOICE AUTOMATED PD SET WITH CASSETTE

MDR report key: 3855198 · Received June 9, 2014

Report

Report Number
1416980-2014-18316
Event Type
Malfunction
Date Received
June 9, 2014
Date of Event
May 17, 2014
Report Date
May 17, 2014
Manufacturer
BAXTER HEALTHCARE - MOUNTAIN HOME
Product Code
FKX
PMA / PMN Number
K102936
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE PATIENT REPORTED THAT THEY HAD NOT OPENED THE CLAMP ON THE PATIENT LINE DURING PRIMING. USE ERRORS AND PROPER USER INSTRUCTIONS ARE ADDRESSED IN ¿THE HOMECHOICE AND HOMECHOICE PRO APD SYSTEMS PATIENT AT-HOME GUIDE¿, WHICH IS SHIPPED WITH EVERY HOMECHOICE DEVICE. THE GUIDE INFORMS THE USER TO OPEN CLAMPS ON LINES CONNECTED TO SOLUTION BAGS AND THE PATIENT LINE TO ENSURE PROPER PRIMING. THE GUIDE INSTRUCTS THE USER TO CHECK THE LINES FOR CLOSED CLAMPS. A REVIEW OF THE LABEL FOR THE PRODUCT FAMILY WILL BE CONDUCTED. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

DURING TROUBLESHOOTING FOR A CHECK PATIENT LINE ALARM, THE HOME PATIENT (HP) STATED THAT THEY HAD NOT OPENED THE CLAMP ON THE PATIENT LINE DURING PRIMING. THE TECHNICAL SERVICE REPRESENTATIVE (TSR) ASSISTED THE PATIENT IN ENDING THERAPY AND ADVISED THE HP TO RESTART THERAPY USING ALL NEW SUPPLIES. THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION REPORTED. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
335132 HOMECHOICE AUTOMATED PD SET WITH CASSETTE SYSTEM, PERITONEAL, AUTOMATIC DELIVERY FKX BAXTER HEALTHCARE - MOUNTAIN HOME

Patients

Seq Age Sex Outcome Treatment
1 64 YR HOMECHOICE