FDA Adverse Event Malfunction Summary report: N

ANIMAS INSULIN INFUSION PUMP

MDR report key: 3855033 · Received June 9, 2014

Report

Report Number
2531779-2014-16174
Event Type
Malfunction
Date Received
June 9, 2014
Report Date
May 22, 2014
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K042873
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

FOLLOW UP #1 SUBMITTED: 07/16/2014. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE. DEVICE EVALUATION: ON EXAMINATION, THE KEYPAD WAS INTACT WITHOUT DAMAGE. TESTING OF THE KEYPAD BUTTON FUNCTION WAS NOT POSSIBLE DUE TO AN UNRELATED ISSUE. THE KEYPAD COVER WAS REMOVED FOR INVESTIGATION AND REVEALED CONTAMINATION ON THE CONTACTS OF THE KEYPAD BUTTONS.

Additional Manufacturer Narrative · 1

(B)(6). THE PUMP NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.

Description of Event or Problem · 1

THE DISTRIBUTOR CONTACTED ANIMAS ON (B)(6) 2014 ALLEGING THE OK KEYPAD BUTTON WAS UNRESPONSIVE. THERE WAS NO REPORTED IMPACT TO THE PATIENT ASSOCIATED WITH THIS COMPLAINT. THIS COMPLAINT IS BEING REPORTED BECAUSE THE ALLEGED KEYPAD BUTTON ISSUE REMAINED UNRESOLVED WITH TROUBLESHOOTING.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
335520 ANIMAS INSULIN INFUSION PUMP INSULIN INFUSION PUMP LZG ANIMAS CORPORATION

Patients

Seq Age Sex Outcome Treatment
1