FDA Adverse Event Malfunction Summary report: N

ANIMAS INSULIN INFUSION PUMP

MDR report key: 3855013 · Received June 9, 2014

Report

Report Number
2531779-2014-16107
Event Type
Malfunction
Date Received
June 9, 2014
Report Date
June 2, 2014
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K042873
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2014 WITH THE FOLLOWING FINDINGS: EVALUATION REVEALED THE KEYPAD IS TORN OVER THE OK BUTTON. A DAMAGED KEYPAD WILL ALLOW CONTAMINATION TO PERMEATE THE BUTTON CONTACTS NEGATIVELY IMPACTING THE BUTTON FUNCTION. THE DAMAGED KEYPAD IS OBVIOUS AND DETECTABLE AND SHOULD ALERT THE USER TO DISCONTINUE USE OF THE PUMP. THE DOWN BUTTON HAS AN INTERMITTENT RESPONSE. THE OK AND UP BUTTONS ARE UNRESPONSIVE. THE CONTRAST BUTTON RESPONDS PROPERLY. REMOVED THE KEYPAD AND FOUND THE OK BUTTON CONTACT INVERTED. CONTAMINATION WAS FOUND UNDER ALL OF THE BUTTON CONTACTS. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE. (B)(6).

Description of Event or Problem · 1

THE PUMP WAS RETURNED FOR INVESTIGATION. EVALUATION REVEALED CONTAMINATION UNDER BUTTONS. THIS REPORT IS MADE BASED ON RESULTS OF INVESTIGATION COMPLETED ON (B)(4) 2014.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
334743 ANIMAS INSULIN INFUSION PUMP INSULIN INFUSION PUMP LZG ANIMAS CORPORATION

Patients

Seq Age Sex Outcome Treatment
1 25 YR