FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 3854967 · Received June 9, 2014

Report

Report Number
2531779-2014-16102
Event Type
Malfunction
Date Received
June 9, 2014
Report Date
May 22, 2014
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K032257
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(6) 2014 WITH THE FOLLOWING FINDINGS: THE PUMP POWERS WITH RETURNED BATTERY CAP. BATTERY CAP IS UNABLE TO FULLY TIGHTEN. BATTERY COMPARTMENT THREADS DAMAGED. A POWER LOSS WAS NOT OBSERVED. BATTERY CAP IS ABLE TO BE REMOVED WITHOUT DIFFICULTY. BATTER CAP THREADS INTACT. BATTERY CAP IS ABLE TO FULLY TIGHTEN ON A TEST PUMP. NO DAMAGE TO BATTERY CAP OBSERVED. (B)(6).

Description of Event or Problem · 1

THE PUMP WAS RETURNED FOR INVESTIGATION. EVALUATION REVEALED BATTERY COMPARTMENT THREADS ARE DAMAGED. THIS REPORT IS MADE BASED ON RESULTS OF INVESTIGATION COMPLETED ON (B)(6) 2014.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
336567 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION

Patients

Seq Age Sex Outcome Treatment
1 32 YR