VENTED AUTOFEED CHAMBER
Report
- Report Number
- 9611451-2014-00493
- Event Type
- Malfunction
- Date Received
- June 9, 2014
- Report Date
- May 8, 2014
- Manufacturer
- FISHER & PAYKEL HEALTHCARE LTD
- Product Code
- BTT
- PMA / PMN Number
- K934140
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CH
- Reporter Occupation
- NURSE
Narratives
(B)(4). METHOD: TWO MR290V VENTED AUTOFEED HUMIDIFICATION CHAMBERS WERE RETURNED TO FISHER & PAYKE HEALTHCARE IN (B)(4) WHERE THEY WERE VISUALLY INSPECTED. ONE OF THE RETURNED DEVICES WAS FROM LOT 130518 (DATE OF MANUFACTURE: 18 MAY 2013) AND THE OTHER WAS FROM LOT 130715 (DATE OF MANUFACTURE: 15 JULY 2013). RESULTS: VISUAL INSPECTION OF THE CHAMBER FROM LOT 130715 REVEALED A CRACK UNDERNEATH THE BAFFLE. THE CRACK SHOWED STRESS MARKS AND SCRATCHES WERE NOTED ON THE SIDE OF THE CHAMBER DOME. ADDITIONAL STRESS MARKS WERE NOTED BETWEEN CHAMBER PORT AND HINGE BRACKET. VISUAL INSPECTION OF THE OTHER CHAMBER REVEALED DENTS ON THE BASE. A VERTICAL CRACK WAS ALSO OBSERVED UNDER THE CHAMBER PORTS. A LOT CHECK REVEALED NO OTHER COMPLAINTS FOR LOT 130518 AND LOT 130715. CONCLUSION: BASED ON THE ABOVE INSPECTION THE DAMAGE OBSERVED ON THE RETURNED CHAMBERS WAS MOST LIKELY CAUSED BY SOME FORM OF IMPACT, POSSIBLY DURING TRANSPORTATION OR STORAGE OF THE CHAMBERS. EVERY MR290 CHAMBER IS PRESSURE TESTED FOLLOWING THE MANUFACTURING PROCESS TO CHECK FOR ANY LEAKS PRESENT IN THE CHAMBER DUE TO CRACKS AND OTHER CAUSES. ANY CHAMBER WHICH FAILS THIS TEST IS REJECTED. IN ADDITION, THE PRESSURE TEST IS FOLLOWED BY A VISUAL INSPECTION OF EACH CHAMBER. THIS SUGGESTS THE RETURNED CHAMBERS WERE DAMAGED AFTER THEY WERE RELEASED FOR DISTRIBUTION. THE USER INSTRUCTIONS THAT ACCOMPANY THE MR290V VENTED AUTOFEED HUMIDIFICATION CHAMBER STATE: "DO NOT USE THE CHAMBER IF THE SEALS ARE NOT INTACT WHEN RECEIVED, OR IF IT HAS BEEN DROPPED." "PERFORM A PRESSURE AND LEAK TEST ON THE BREATHING SYSTEM AND CHECK FOR OCCLUSIONS BEFORE CONNECTING TO A PATIENT." "SET APPROPRIATE VENTILATOR ALARMS."
(B)(4). THE COMPLAINT DEVICES ARE CURRENTLY EN ROUTE TO FISHER & PAYKEL HEALTHCARE IN (B)(4) FOR EVALUATION. WE WILL PROVIDE A FOLLOW UP REPORT UPON COMPLETION OF OUR INVESTIGATION.
A HOSPITAL IN (B)(6) REPORTED VIA A DISTRIBUTOR THAT TWO MR290 AUTOFEED HUMIDIFICATION CHAMBERS WERE LEAKING WHEN CONNECTED TO A VENTILATOR. THIS WAS FOUND PRIOR TO PATIENT USE.
A HOSPITAL IN (B)(6) REPORTED VIA A DISTRIBUTOR THAT TWO MR290 AUTOFEED HUMIDIFICATION CHAMBERS WERE LEAKING WHEN CONNECTED TO A VENTILATOR. THIS WAS FOUND PRIOR TO PATIENT USE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 336537 | VENTED AUTOFEED CHAMBER | AUTOFEED CHAMBER | BTT | FISHER & PAYKEL HEALTHCARE LTD | MR290HFV | 130715 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | FPH RT235 INFANT BREATHING CIRCUIT| FPH RT235 INFANT BREATHING CIRCUIT |