ONETOUCHPING GLUCOSEMGMTSYSTEM
Report
- Report Number
- 2531779-2014-16104
- Event Type
- Malfunction
- Date Received
- June 9, 2014
- Report Date
- May 23, 2014
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- PATIENT
Narratives
THE PUMP WAS RETURNED FOR INVESTIGATION AND DEVICE EVALUATION WAS COMPLETED BY PRODUCT ANALYSIS ON 05/23/2014 WITH THE FOLLOWING FINDINGS: NO POWER ON RESET EVENTS WERE RECORDING IN THE BLACK BOX TO SUPPORT TIME AND DATE RESETS. PUMP WAS EXERCISED FOR 24 HOURS, THEN BATTERY WAS REMOVED FOR 6 HOURS, AND THE TIME AND DATE RESET TO DEFAULT SETTINGS. PUMP WAS OPEN TO INVESTIGATE AND THE INTERNAL BATTERY WAS FOUND TO BE LEAKING AND UNABLE TO HOLD A CHARGE. UNRELATED TO THE COMPLAINT, THERE IS A CRACK IN THE BATTERY COMPARTMENT AND THE DISPLAY IS DIM AND PINK. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE.
THERE WAS NO ADVERSE EVENT ASSOCIATED WITH THIS COMPLAINT. THE PUMP WAS RETURNED FOR PRODUCT ANALYSIS INVESTIGATION AND THE EVALUATION REVEALED A LEAKING INTERNAL BATTERY, A DIM PINK DISPLAY AND A CRACKED BATTERY COMPARTMENT. THIS REPORT IS MADE BASED ON THE RESULTS OF AN INVESTIGATION COMPLETED ON 05/23/2014.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 336547 | ONETOUCHPING GLUCOSEMGMTSYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 65 YR |