FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 3854937 · Received June 9, 2014

Report

Report Number
2531779-2014-16104
Event Type
Malfunction
Date Received
June 9, 2014
Report Date
May 23, 2014
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

THE PUMP WAS RETURNED FOR INVESTIGATION AND DEVICE EVALUATION WAS COMPLETED BY PRODUCT ANALYSIS ON 05/23/2014 WITH THE FOLLOWING FINDINGS: NO POWER ON RESET EVENTS WERE RECORDING IN THE BLACK BOX TO SUPPORT TIME AND DATE RESETS. PUMP WAS EXERCISED FOR 24 HOURS, THEN BATTERY WAS REMOVED FOR 6 HOURS, AND THE TIME AND DATE RESET TO DEFAULT SETTINGS. PUMP WAS OPEN TO INVESTIGATE AND THE INTERNAL BATTERY WAS FOUND TO BE LEAKING AND UNABLE TO HOLD A CHARGE. UNRELATED TO THE COMPLAINT, THERE IS A CRACK IN THE BATTERY COMPARTMENT AND THE DISPLAY IS DIM AND PINK. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE.

Description of Event or Problem · 1

THERE WAS NO ADVERSE EVENT ASSOCIATED WITH THIS COMPLAINT. THE PUMP WAS RETURNED FOR PRODUCT ANALYSIS INVESTIGATION AND THE EVALUATION REVEALED A LEAKING INTERNAL BATTERY, A DIM PINK DISPLAY AND A CRACKED BATTERY COMPARTMENT. THIS REPORT IS MADE BASED ON THE RESULTS OF AN INVESTIGATION COMPLETED ON 05/23/2014.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
336547 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION

Patients

Seq Age Sex Outcome Treatment
1 65 YR