FDA Adverse Event Malfunction Summary report: N

MINICAP TRANSFER SET

MDR report key: 3854768 · Received June 6, 2014

Report

Report Number
1416980-2014-18287
Event Type
Malfunction
Date Received
June 6, 2014
Date of Event
May 14, 2014
Report Date
May 14, 2014
Manufacturer
BAXTER HEALTHCARE - MOUNTAIN HOME
Product Code
KDJ
PMA / PMN Number
K882498
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TW
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

(B)(4). A REQUEST FOR THE RETURN OF THE DEVICE HAS BEEN MADE. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Additional Manufacturer Narrative · 1

(B)(4). A REVIEW OF ALL BATCH RECORD DOCUMENTS WAS CONDUCTED WITH NO ISSUES NOTED DURING THE MANUFACTURING PROCESS. THERE WERE NO DEVIATIONS FROM STANDARD PROCEDURE AND NO EXCEPTIONS RELATED TO THE REPORTED CONDITION WERE NOTED. A COMPANION SAMPLE WAS RECEIVED FOR EVALUATION. A VISUAL INSPECTION WAS PERFORMED WITH LOOSE AND EMBEDDED MATRICULATE MATTER NOTED. THE REPORTED PROBLEM WAS IDENTIFIED BUT THE CAUSE COULD NOT BE DETERMINED.

Description of Event or Problem · 1

IT WAS REPORTED THAT A BLACK SPOT WAS OBSERVED ON THE TUBING OF A MINICAP EXTEND LIFE PD TRANSFER SET. THIS WAS NOTICED BEFORE PATIENT USE, AS THE CARTON WAS BEING OPENED. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE. THIS IS REPORT 1 OF 2.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
331254 MINICAP TRANSFER SET SET, ADMINISTRATION, FOR PERITONEAL DIALYSIS, DISPOSABLE KDJ BAXTER HEALTHCARE - MOUNTAIN HOME H13I06037

Patients

Seq Age Sex Outcome Treatment
1