FDA Adverse Event Malfunction Summary report: N

QUADRA ASSURA CRT-D, DF4 CONNECTOR

MDR report key: 3854652 · Received June 6, 2014

Report

Report Number
2938836-2014-11613
Event Type
Malfunction
Date Received
June 6, 2014
Date of Event
April 1, 2014
Manufacturer
ST. JUDE MEDICAL, INC., CRMD
Product Code
NIK
PMA / PMN Number
P910023
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UT, US
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 1

ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED.

Description of Event or Problem · 1

IT WAS REPORTED THAT A MERLIN.NET TRANSMISSION SHOWED EPISODES OF NON-SUSTAINED LEAD NOISE DUE TO POST-PACED T-WAVE OVERSENSING. THE OVERSENSING WAS OBSERVED ON THE EGMS. THE PATIENT WAS ASYMPTOMATIC. PROGRAMMING CHANGES WERE RECOMMENDED AND THE DEVICE REMAINS IMPLANTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
332154 QUADRA ASSURA CRT-D, DF4 CONNECTOR IMPLANTABLE CARDIOVERTER DEFIBRILLATOR NIK ST. JUDE MEDICAL, INC., CRMD CD3265-40Q NA

Patients

Seq Age Sex Outcome Treatment
1 61 YR