FDA Adverse Event
Malfunction
Summary report: N
QUADRA ASSURA CRT-D, DF4 CONNECTOR
MDR report key: 3854652
·
Received June 6, 2014
Report
- Report Number
- 2938836-2014-11613
- Event Type
- Malfunction
- Date Received
- June 6, 2014
- Date of Event
- April 1, 2014
- Manufacturer
- ST. JUDE MEDICAL, INC., CRMD
- Product Code
- NIK
- PMA / PMN Number
- P910023
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- UT, US
- Reporter Occupation
- NURSE
Narratives
Additional Manufacturer Narrative · 1
ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED.
Description of Event or Problem · 1
IT WAS REPORTED THAT A MERLIN.NET TRANSMISSION SHOWED EPISODES OF NON-SUSTAINED LEAD NOISE DUE TO POST-PACED T-WAVE OVERSENSING. THE OVERSENSING WAS OBSERVED ON THE EGMS. THE PATIENT WAS ASYMPTOMATIC. PROGRAMMING CHANGES WERE RECOMMENDED AND THE DEVICE REMAINS IMPLANTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 332154 | QUADRA ASSURA CRT-D, DF4 CONNECTOR | IMPLANTABLE CARDIOVERTER DEFIBRILLATOR | NIK | ST. JUDE MEDICAL, INC., CRMD | CD3265-40Q | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 61 YR |