FDA Adverse Event Malfunction Summary report: N

HOMECHOICE PRO

MDR report key: 3854646 · Received June 6, 2014

Report

Report Number
1416980-2014-18283
Event Type
Malfunction
Date Received
June 6, 2014
Date of Event
April 4, 2014
Report Date
May 14, 2014
Manufacturer
BAXTER HEALTHCARE - LARGO
Product Code
FKX
PMA / PMN Number
K102936
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). EVALUATION SUMMARY: THE DEVICE WAS RECEIVED AND THE EVALUATION IS COMPLETE. THIS IS AN ANCILLARY SERVICE EVENT. VISUAL INSPECTION AND FUNCTIONAL TESTING WERE PERFORMED. THE INCREASED INTRA-PERITONEAL VOLUME (IIPV) EVENT WAS IDENTIFIED THROUGH AN EVENT HISTORY LOG REVIEW. UPON CONCLUSION OF THE INVESTIGATION, THE CAUSE WAS NOT ABLE TO BE DETERMINED. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

DURING EVALUATION OF A RETURNED HOMECHOICE DEVICE, ONE INCREASED INTRA-PERITONEAL VOLUME (IIPV) EVENT WAS IDENTIFIED. THIS EVENT OCCURRED DURING THE THERAPY INITIATED ON (B)(6) 2014 AT 21:05:43. DURING NIGHT DRAIN CYCLE FIVE, THE PATIENT'S ULTRAFILTRATION READING WAS 1760ML, INDICATING THE HOME PATIENT DRAINED 1540ML MORE THAN THEIR MAXIMUM PROGRAMMED FILL VOLUME OF 2200ML. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
332152 HOMECHOICE PRO SYSTEM, PERITONEAL, AUTOMATIC DELIVERY FKX BAXTER HEALTHCARE - LARGO

Patients

Seq Age Sex Outcome Treatment
1