HOMECHOICE AUTOMATED PD SET WITH CASSETTE
Report
- Report Number
- 1416980-2014-18282
- Event Type
- Malfunction
- Date Received
- June 6, 2014
- Date of Event
- May 14, 2014
- Report Date
- May 14, 2014
- Manufacturer
- BAXTER HEALTHCARE - MOUNTAIN HOME
- Product Code
- FKX
- PMA / PMN Number
- K102936
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- DE, US
- Reporter Occupation
- OTHER
Narratives
(B)(4). THE DEVICE WAS NOT RETURNED AND THE LOT NUMBER WAS UNKNOWN; THEREFORE, A DEVICE ANALYSIS COULD NOT BE COMPLETED. HOWEVER, A CLAMP BEING LEFT OPEN ON UNUSED LINES IS A KNOWN CAUSE OF THIS ALARM. PER ¿THE HOMECHOICE AND HOMECHOICE PRO APD SYSTEMS PATIENT AT-HOME GUIDE¿, USERS ARE INSTRUCTED THAT CLAMPS ON ANY UNUSED SOLUTION LINES MUST REMAIN CLOSED WHEN PERFORMING PERITONEAL DIALYSIS THERAPY. A FORMAL REVIEW OF THE LABEL FOR THE PRODUCT FAMILY WILL BE CONDUCTED. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A FOLLOW-UP REPORT WILL BE SUBMITTED.
IT WAS REPORTED THAT A SYSTEM ERROR (SE) 2240 (AIR IN SET) ALARM OCCURRED DURING DWELL THREE OF FOUR ON THE HOME CHOICE (HC). THE HOME PATIENT (HP) WAS CONNECTED AT THE TIME OF THE EVENT. THE CALLER INDICATED THAT A CLAMP WAS LEFT OPENED ON AN UNUSED SUPPLY LINE. THE TECHNICAL SERVICE REPRESENTATIVE (TSR) ASSISTED THE HP TO CYCLE POWER AND DISCONNECTED. THE HP INDICATED THEY WOULD COMPLETE THERAPY WITH MANUAL SUPPLIES. NO PATIENT INJURY OR MEDICAL INTERVENTION WAS INDICATED AS A RESULT OF THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 332489 | HOMECHOICE AUTOMATED PD SET WITH CASSETTE | SYSTEM, PERITONEAL, AUTOMATIC DELIVERY | FKX | BAXTER HEALTHCARE - MOUNTAIN HOME |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 54 YR | HOMECHOICE |