PROMUS ELEMENT ?
Report
- Report Number
- 2134265-2014-03225
- Event Type
- Malfunction
- Date Received
- June 6, 2014
- Date of Event
- April 24, 2014
- Report Date
- May 14, 2014
- Manufacturer
- BOSTON SCIENTIFIC - GALWAY
- Product Code
- NIQ
- PMA / PMN Number
- SIMILAR
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CN
- Reporter Occupation
- PHYSICIAN
Narratives
DEVICE EVALUATED BY MFR.: A VISUAL EXAMINATION OF THE CRIMPED STENT FOUND STENT STRUTS ON THE PROXIMAL END OF THE CRIMPED STENT WERE DISTORTED AND LIFTED UPWARDS FROM THE STENT PROFILE. THE BUMPER TIP OF THE DEVICE SHOWED NO SIGNS OF DAMAGE. THE BALLOON CONES PROFILES WERE REVIEWED AND NO ISSUES WERE NOTED WITH THE OVERALL BALLOON PROFILE. THE BALLOON WINGS WERE TIGHTLY WRAPPED AND EVENLY FOLDED AND WERE NOT SUBJECTED TO POSITIVE PRESSURE. A VISUAL AND TACTILE EXAMINATION FOUND MULTIPLE KINKS ALONG THE HYPOTUBE SHAFT. A VISUAL AND TACTILE EXAMINATION FOUND NO ISSUES WITH THE PROFILE. THE MANUFACTURING BATCH RECORD REVIEW CONFIRMED THAT THE DEVICE MET ALL MATERIAL, ASSEMBLY AND PERFORMANCE SPECIFICATIONS. THE MOST PROBABLE ROOT CAUSE IS OPERATIONAL CONTEXT AS DEVICE PERFORMANCE WAS LIMITED DUE TO ANATOMICAL PROCEDURAL FACTORS. (B)(4).
DEVICE IS A COMBINATION PRODUCT. (B)(4).
IT WAS REPORTED THAT STENT DAMAGE OCCURRED. THE 90% STENOSED, 26X3.5MM, ECCENTRIC AND DE NOVO TARGET LESION WAS LOCATED IN THE MILDLY CALCIFIED AND MILDLY TORTUOUS RIGHT CORONARY ARTERY (RCA). PREDILATION WAS PERFORMED USING AN UNSPECIFIED BALLOON CATHETER. A 3.50X28MM PROMUS ELEMENT ¿ DRUG-ELUTING STENT WAS THEN SELECTED TO TREAT THE TARGET LESION BUT WAS UNABLE TO CROSS. IT WAS NOTICED THAT STENT BODY WAS LIFTED. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED AND THE PATIENT'S STATUS WAS STABLE.
IT WAS REPORTED THAT STENT DAMAGE OCCURRED. THE 90% STENOSED, 26X3.5MM, ECCENTRIC AND DE NOVO TARGET LESION WAS LOCATED IN THE MILDLY CALCIFIED AND MILDLY TORTUOUS RIGHT CORONARY ARTERY (RCA). PREDILATION WAS PERFORMED USING AN UNSPECIFIED BALLOON CATHETER. A 3.50X28MM PROMUS ELEMENT ¿ DRUG-ELUTING STENT WAS THEN SELECTED TO TREAT THE TARGET LESION BUT WAS UNABLE TO CROSS. IT WAS NOTICED THAT STENT BODY WAS LIFTED. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED AND THE PATIENT'S STATUS WAS STABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 331160 | PROMUS ELEMENT ? | STENT, CORONARY, DRUG-ELUTING | NIQ | BOSTON SCIENTIFIC - GALWAY | H7493911328350 | 16067778 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 45 YR |