RESOLUTION CLIP CLIPPING DEVICE
Report
- Report Number
- 3005099803-2014-02099
- Event Type
- Malfunction
- Date Received
- June 6, 2014
- Date of Event
- May 16, 2014
- Report Date
- May 16, 2014
- Manufacturer
- BOSTON SCIENTIFIC - MARLBOROUGH
- Product Code
- KOG
- PMA / PMN Number
- K040148
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CN
- Reporter Occupation
- PHYSICIAN
Narratives
REPORTED EVENT OF CLIP FAILED TO RELEASE FROM THE CATHETER. ALTHOUGH THE SUSPECT DEVICE HAS BEEN RECEIVED, THE EVALUATION HAS NOT BEEN COMPLETED. THEREFORE, THE CAUSE OF THE REPORTED MALFUNCTION HAS NOT BEEN DETERMINED. UPON COMPLETION OF THE FAILURE ANALYSIS OF THE COMPLAINT DEVICE, IF THERE IS ANY FURTHER RELEVANT INFORMATION FROM THAT REVIEW, A SUPPLEMENTAL MDR WILL BE FILED.
A VISUAL EXAMINATION OF THE RETURNED DEVICE NOTED THAT THE CLIP ASSEMBLY WAS FULLY DEPLOYED AND NOT RETURNED. IT WAS ALSO NOTICED THAT THE COIL WAS BENT/KINKED AT THE DISTAL END. THERE WAS A KINK IN THE CONTROL WIRE. BASED ON THE CONDITION OF THE RETURNED DEVICE, THE REPORTED FAILURE COULD NOT BE CONFIRMED. HOWEVER, IT IS LIKELY THAT THE DIFFICULTIES EXPERIENCED DURING DEPLOYMENT WERE DUE TO ANATOMICAL OR PROCEDURAL FACTORS SUCH AS TORTUOUS ANATOMY OR MANEUVERING OF THE DEVICE. THEREFORE, A REVIEW AND ANALYSIS OF ALL AVAILABLE INFORMATION INDICATED THAT THE MOST PROBABLE ROOT CAUSE IS OPERATIONAL CONTEXT. A REVIEW OF THE DEVICE HISTORY RECORD (DHR) WAS PERFORMED AND CONFIRMED THAT THIS DEVICE MET ALL MATERIAL, ASSEMBLY AND PERFORMANCE SPECIFICATIONS. A SEARCH OF THE COMPLAINT DATABASE REVEALED THAT NO OTHER COMPLAINTS EXIST FOR THE SPECIFIED LOT.
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A RESOLUTION CLIP DEVICE WAS USED IN THE LOWER MIDDLE SEGMENT OF THE ESOPHAGUS DURING A POLYP SNARE PROCEDURE PERFORMED ON (B)(6) 2014. ACCORDING TO THE COMPLAINANT, DURING THE PROCEDURE, THE CLIP GRASPED AND LOCKED ONTO TISSUE; HOWEVER THE CLIP WOULD NOT RELEASE FROM THE CATHETER TO DEPLOY. THE CLIP HAD TO BE PULLED OFF FROM THE TISSUE. WHEN THE DEVICE REMOVED, IT WAS DISCOVERED THAT THE CLIP HAD DETACHED FROM THE CATHETER OUTSIDE THE PATIENT. ANOTHER RESOLUTION CLIP WAS USED TO COMPLETE THE PROCEDURE. NO PATIENT COMPLICATIONS RESULTED FROM THIS EVENT. THE PATIENT'S CONDITION FOLLOWING THE PROCEDURE WAS REPORTED TO BE STABLE.
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A RESOLUTION CLIP DEVICE WAS USED IN THE LOWER MIDDLE SEGMENT OF THE ESOPHAGUS DURING A POLYP SNARE PROCEDURE PERFORMED ON (B)(6) 2014. ACCORDING TO THE COMPLAINANT, DURING THE PROCEDURE, THE CLIP GRASPED AND LOCKED ONTO TISSUE; HOWEVER THE CLIP WOULD NOT RELEASE FROM THE CATHETER TO DEPLOY. THE CLIP HAD TO BE PULLED OFF FROM THE TISSUE. WHEN THE DEVICE REMOVED, IT WAS DISCOVERED THAT THE CLIP HAD DETACHED FROM THE CATHETER OUTSIDE THE PATIENT. ANOTHER RESOLUTION CLIP WAS USED TO COMPLETE THE PROCEDURE. NO PATIENT COMPLICATIONS RESULTED FROM THIS EVENT. THE PATIENT'S CONDITION FOLLOWING THE PROCEDURE WAS REPORTED TO BE STABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 332775 | RESOLUTION CLIP CLIPPING DEVICE | ENDOSCOPE AND/OR ACCESSORIES | KOG | BOSTON SCIENTIFIC - MARLBOROUGH | M00522610 | ML000649C3 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 54 YR |