FDA Adverse Event Malfunction Summary report: N

MINICAP TRANSFER SET

MDR report key: 3853852 · Received June 6, 2014

Report

Report Number
1416980-2014-18177
Event Type
Malfunction
Date Received
June 6, 2014
Date of Event
May 14, 2014
Report Date
May 14, 2014
Manufacturer
BAXTER HEALTHCARE - MOUNTAIN HOME
Product Code
KDJ
PMA / PMN Number
K882498
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TW
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). A PHOTOGRAPH OF THE DEVICE WAS RECEIVED FOR EVALUATION. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Additional Manufacturer Narrative · 1

(B)(4). A REVIEW OF ALL BATCH RECORD DOCUMENTS WAS CONDUCTED WITH NO ISSUES NOTED DURING THE MANUFACTURING PROCESS. THERE WERE NO DEVIATIONS FROM STANDARD PROCEDURE AND NO EXCEPTIONS RELATED TO THE REPORTED CONDITION WERE NOTED. A SAMPLE WAS RECEIVED FOR EVALUATION. A VISUAL INSPECTION WAS PERFORMED WITH NO ISSUES NOTED. A LEAK TEST, CLAMP FUNCTION TEST, AND CLAMP FUNCTION TEST WERE PERFORMED WITH NO ISSUES NOTED. AN INTEGRITY OF SEAL TEST WAS PERFORMED WITH NO LEAK NOTED. THE INTERIOR DIAMETER OF THE PATIENT CONNECTOR WAS MEASURED AND WAS FOUND TO BE WITHIN SPECIFICATION. THE REPORTED PROBLEM COULD NOT BE IDENTIFIED OR DUPLICATED VIA AN EVALUATION. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A FOLLOW-UP REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

IT WAS REPORTED THAT A MINICAP TRANSFER SET COULD NOT BE COMPLETED LOCKED, OR SECURELY CONNECTED, WITH A TITANIUM ADAPTOR. THIS OCCURRED DURING PATIENT USE. THE PATIENT STATED THAT THE ISSUE WAS RESOLVED WHEN THE TRANSFER SET WAS REPLACED. THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
332569 MINICAP TRANSFER SET SET, ADMINISTRATION, FOR PERITONEAL DIALYSIS, DISPOSABLE KDJ BAXTER HEALTHCARE - MOUNTAIN HOME H13J29038

Patients

Seq Age Sex Outcome Treatment
1 TITANIUM ADAPTOR