FDA Adverse Event Malfunction Summary report: N

SYSTEM, PERITONEAL, AUTOMATIC DELIVERY

MDR report key: 3853820 · Received June 6, 2014

Report

Report Number
1416980-2014-18174
Event Type
Malfunction
Date Received
June 6, 2014
Date of Event
May 14, 2014
Report Date
May 14, 2014
Manufacturer
BAXTER HEALTHCARE - MOUNTAIN HOME
Product Code
FKX
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
KY, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). AN ALARM INDICATIVE OF A POTENTIAL MALFUNCTION OF THE DISPOSABLE CASSETTE WAS IDENTIFIED.  AS THE CASSETTE WAS NOT RETURNED AND THE LOT NUMBER IS UNKNOWN, A DEVICE ANALYSIS CANNOT BE COMPLETED. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

IT WAS REPORTED THAT A SYSTEM ERROR (SE) 2240 (AIR IN SET/LINE) ALARM OCCURRED DURING THE INITIAL DRAIN ON THE HOME CHOICE (HC). THE HOME PATIENT (HP) WAS CONNECTED AT THE TIME OF THE EVENT. THERE WAS NOTHING FOUND DURING TROUBLESHOOTING THAT WOULD CAUSE OR CONTRIBUTE TO THE ALARM. THE TECHNICAL SERVICE REPRESENTATIVE (TSR) ADVISED THE HP TO CLOSE THE CLAMPS AND CYCLED POWER. THE TSR ADVISED THE HP TO START OVER WITH NEW SUPPLIES. NO PATIENT INJURY OR MEDICAL INTERVENTION WAS INDICATED AS A RESULT OF THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
332708 SYSTEM, PERITONEAL, AUTOMATIC DELIVERY FKX BAXTER HEALTHCARE - MOUNTAIN HOME

Patients

Seq Age Sex Outcome Treatment
1 85 YR HOMECHOICE