FDA Adverse Event Malfunction Summary report: N

SYSTEM, PERITONEAL, AUTOMATIC DELIVERY

MDR report key: 3853688 · Received June 6, 2014

Report

Report Number
1416980-2014-18160
Event Type
Malfunction
Date Received
June 6, 2014
Date of Event
May 12, 2014
Report Date
May 12, 2014
Manufacturer
BAXTER HEALTHCARE - MOUNTAIN HOME
Product Code
FKX
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE CLAMP ON THE PATIENT LINE NOT BEING OPENED BEFORE THE PATIENT WAS CONNECTED IS A KNOWN CAUSE OF THIS ALARM. USE ERRORS AND PROPER USER INSTRUCTIONS ARE ADDRESSED IN ¿THE HOMECHOICE AND HOMECHOICE PRO APD SYSTEMS PATIENT AT-HOME GUIDE¿, WHICH IS SHIPPED WITH EVERY HOMECHOICE DEVICE. IT INFORMS THE USER TO OPEN CLAMPS ON LINES CONNECTED TO SOLUTION BAGS AND THE PATIENT LINE TO ENSURE PROPER PRIMING. IT INSTRUCTS THE USER TO CHECK THE LINES FOR CLOSED CLAMPS. A REVIEW OF THE LABEL FOR THE PRODUCT FAMILY WILL BE CONDUCTED. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

IT WAS REPORTED THAT A SYSTEM ERROR 2240 (AIR IN SET/LINE) ALARM OCCURRED DURING INITIAL DRAIN ON THE HOMECHOICE. THE PATIENT WAS CONNECTED AT THE TIME OF THE ALARM. DURING TROUBLESHOOTING, THE CAREGIVER (CG) STATED THAT THE HOME PATIENT WAS CONNECTED WHILE THE PATIENT LINE CLAMP WAS CLOSED FOR PRIMING. THE TECHNICAL SERVICE REPRESENTATIVE (TSR) REVIEWED PROPER PROCEDURES WITH THE CG AND ASSISTED THE CG IN ENDING THERAPY. THE CG PLANNED TO RESTART THERAPY USING ALL NEW SUPPLIES. THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION REPORTED. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
332983 SYSTEM, PERITONEAL, AUTOMATIC DELIVERY FKX BAXTER HEALTHCARE - MOUNTAIN HOME

Patients

Seq Age Sex Outcome Treatment
1 65 YR HOMECHOICE