FDA Adverse Event
Malfunction
Summary report: N
840 VENTILATOR
MDR report key: 3853599
·
Received April 3, 2014
Report
- Report Number
- 8020893-2014-00781
- Event Type
- Malfunction
- Date Received
- April 3, 2014
- Report Date
- March 5, 2014
- Manufacturer
- COVIDIEN, FORMERLY NELLCOR
- Product Code
- CBK
- PMA / PMN Number
- K970460
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MS, US
- Reporter Occupation
- BIOMEDICAL ENGINEER
Narratives
Additional Manufacturer Narrative · 1
THE COVIDIEN CUSTOMER SUPPORT ENGINEER (CSE) INSPECTED THE DEVICE AND VERIFIED THE MALFUNCTION. THE CSE REPLACED THE BACKLIGHT INVERTER PRINTED CIRCUIT BOARD (PCB). THE DEVICE PASSED ALL CALIBRATIONS, SHORT SELF-TEST, EXTENDED SELF-TEST (EST) AND PERFORMANCE VERIFICATION TEST ACCORDING TO MANUFACTURER SPECIFICATIONS. (B)(4).
Description of Event or Problem · 1
A REPORT RECEIVED FROM THE USA STATED AN 840 VENTILATOR EXPERIENCED A BLANK LOWER DISPLAY. THE VENTILATOR WAS NOT BEING USED ON A PATIENT AT THE TIME THE EVENT OCCURRED. THE DATE OF THE EVENT IS UNKNOWN.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 202793 | 840 VENTILATOR | CBK | COVIDIEN, FORMERLY NELLCOR | 840 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |