FDA Adverse Event
Malfunction
Summary report: N
840 VENTILATOR
MDR report key: 3853596
·
Received April 3, 2014
Report
- Report Number
- 8020893-2014-00774
- Event Type
- Malfunction
- Date Received
- April 3, 2014
- Report Date
- March 5, 2014
- Manufacturer
- COVIDIEN, FORMERLY NELLCOR PURITAN
- Product Code
- CBK
- PMA / PMN Number
- K970460
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OH, US
- Reporter Occupation
- BIOMEDICAL ENGINEER
Narratives
Additional Manufacturer Narrative · 1
THE COVIDIEN CUSTOMER SUPPORT ENGINEER (CSE) INSPECTED THE DEVICE AND VERIFIED THE REPORTED ISSUE. THE CSE REPLACED THE GRAPHICAL USER INTERFACE (GUI) TOUCH FRAME, PRINTED CIRCUIT BOARD (PCB). THE DEVICE PASSED ALL CALIBRATIONS, SHORT SELF-TEST, EXTENDED SELF-TEST (EST) AND PERFORMANCE VERIFICATION TEST ACCORDING TO MFR SPECS. (B)(4).
Description of Event or Problem · 1
A REPORT RECEIVED FROM THE USA STATED AN 840 VENTILATOR EXPERIENCED AN UNRESPONSIVE TOUCHSCREEN. THE VENTILATOR WAS NOT IN USE ON A PT AT THE TIME OF THE EVENT. THE DATE OF THE EVENT IS UNK.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 202596 | 840 VENTILATOR | CBK | COVIDIEN, FORMERLY NELLCOR PURITAN | 840 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |