FDA Adverse Event Malfunction Summary report: N

840 VENTILATOR

MDR report key: 3853596 · Received April 3, 2014

Report

Report Number
8020893-2014-00774
Event Type
Malfunction
Date Received
April 3, 2014
Report Date
March 5, 2014
Manufacturer
COVIDIEN, FORMERLY NELLCOR PURITAN
Product Code
CBK
PMA / PMN Number
K970460
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Additional Manufacturer Narrative · 1

THE COVIDIEN CUSTOMER SUPPORT ENGINEER (CSE) INSPECTED THE DEVICE AND VERIFIED THE REPORTED ISSUE. THE CSE REPLACED THE GRAPHICAL USER INTERFACE (GUI) TOUCH FRAME, PRINTED CIRCUIT BOARD (PCB). THE DEVICE PASSED ALL CALIBRATIONS, SHORT SELF-TEST, EXTENDED SELF-TEST (EST) AND PERFORMANCE VERIFICATION TEST ACCORDING TO MFR SPECS. (B)(4).

Description of Event or Problem · 1

A REPORT RECEIVED FROM THE USA STATED AN 840 VENTILATOR EXPERIENCED AN UNRESPONSIVE TOUCHSCREEN. THE VENTILATOR WAS NOT IN USE ON A PT AT THE TIME OF THE EVENT. THE DATE OF THE EVENT IS UNK.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
202596 840 VENTILATOR CBK COVIDIEN, FORMERLY NELLCOR PURITAN 840

Patients

Seq Age Sex Outcome Treatment
1