HOMECHOICE AUTOMATED PD SET WITH CASSETTE
MDR report key: 3853326
·
Received June 6, 2014
Report
- Report Number
- 1416980-2014-18155
- Event Type
- Malfunction
- Date Received
- June 6, 2014
- Date of Event
- May 14, 2014
- Report Date
- May 14, 2014
- Manufacturer
- BAXTER HEALTHCARE - SINGAPORE
- Product Code
- FKX
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SP
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
(B)(4). A REVIEW OF ALL BATCH RECORD DOCUMENTS WAS PERFORMED WITH NO ISSUES NOTED DURING THE MANUFACTURING PROCESS. THERE WERE NO DEVIATIONS FROM STANDARD PROCEDURE AND NO EXCEPTIONS RELATED TO THE REPORTED CONDITION WERE NOTED. AS THE CASSETTE WAS NOT RETURNED A DEVICE ANALYSIS CANNOT BE COMPLETED. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.
Description of Event or Problem · 1
IT WAS REPORTED THAT A SYSTEM ERROR (SE) 2240 (AIR IN SET) ALARM OCCURRED DURING DWELL SEVEN OF SEVEN ON THE HOME CHOICE. NO PATIENT INJURY OR MEDICAL INTERVENTION WAS INDICATED AS A RESULT OF THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.