FDA Adverse Event Injury Summary report: N

SIGMA HP UNI TIB TRAY SZ3 LMRL

MDR report key: 3853276 · Received June 6, 2014

Report

Report Number
1818910-2014-20393
Event Type
Injury
Date Received
June 6, 2014
Date of Event
May 13, 2014
Report Date
May 13, 2014
Manufacturer
1219655 DEPUY-RAYNHAM, A DIV. OF DEPUY ORTHOPAEDIC
Product Code
HRY
PMA / PMN Number
PK070267
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
VT, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THIS COMPLAINT IS STILL UNDER INVESTIGATION. DEPUY WILL NOTIFY THE FDA OF THE RESULTS OF THIS INVESTIGATION ONCE IT HAS BEEN COMPLETED. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Additional Manufacturer Narrative · 1

ADDITIONAL NARRATIVE: THE DEVICES ASSOCIATED WITH THIS REPORT WERE NOT RETURNED. A COMPLAINT DATABASE SEARCH FINDS NO OTHER REPORTED INCIDENTS AGAINST THE PROVIDED PRODUCT AND LOT COMBINATION FOR THE TIBIAL TRAY. REVIEW OF THE DEVICE HISTORY RECORDS AND/OR A LOT SPECIFIC COMPLAINT DATABASE SEARCH WAS NOT POSSIBLE AS THE LOT CODE REQUIRED WAS NOT PROVIDED FOR THE CEMENT. REQUESTS FOR ADDITIONAL INVESTIGATIONAL INPUTS WERE MADE IN ACCORDANCE WITH WI-7915 APPENDIX A. NO ADDITIONAL INFORMATION WAS OBTAINED. THE INVESTIGATION COULD NOT VERIFY OR IDENTIFY ANY PRODUCT CONTRIBUTION TO THE REPORTED EVENT WITH THE INFORMATION PROVIDED. BASED ON THE INABILITY TO IDENTIFY ROOT CAUSE, THE NEED FOR CORRECTIVE ACTION WAS NOT INDICATED. DEPUY CONSIDERS THE INVESTIGATION CLOSED AT THIS TIME. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THE INFORMATION WILL BE REVIEWED AND THE INVESTIGATION WILL BE RE-OPENED AS NECESSARY.

Description of Event or Problem · 1

PATIENT WAS REVISED BECAUSE OF PAIN AND TIBIAL LOOSENING. LOOSENING WAS AT THE BONE/CEMENT AND CEMENT/IMPLANT INTERFACES. CEMENT WAS DEPUY MANUFACTURED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
332173 SIGMA HP UNI TIB TRAY SZ3 LMRL KNEE TIBIAL TRAY HRY 1219655 DEPUY-RAYNHAM, A DIV. OF DEPUY ORTHOPAEDIC 329047

Patients

Seq Age Sex Outcome Treatment
1 66 YR Required Intervention