FDA Adverse Event Malfunction Summary report: N

HOMEPUMP ECLIPSE: 400ML, 100ML

MDR report key: 3853252 · Received April 3, 2014

Report

Report Number
2026095-2014-00038
Event Type
Malfunction
Date Received
April 3, 2014
Date of Event
February 28, 2014
Report Date
March 10, 2014
Manufacturer
I-FLOW, LLC
Product Code
MEB
PMA / PMN Number
K052117
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 1

METHOD: THE DEVICE HAS BEEN RETURNED FOR AN EVALUATION AND INVESTIGATION. A VISUAL INVESTIGATION WAS PERFORMED ON THE RETURNED UNIT AND A REVIEW OF THE DEVICE HISTORY RECORDS (DHR) IS IN PROGRESS FOR THE REPORTED LOT NUMBER. RESULTS: AT THIS TIME THE EVALUATION IS STILL IN PROGRESS. RESULTS WILL BE PROVIDED ONCE COMPLETED. CONCLUSIONS: THE INVESTIGATION AND EVALUATION ARE ONGOING, ONCE COMPLETED A FOLLOW-UP REPORT WILL BE SUBMITTED. INFO FROM THIS INCIDENT WILL BE INCLUDED IN OUR PRODUCT COMPLAINT AND MDR TREND REPORTING SYSTEM. ADDITIONAL INVESTIGATION MAY ARISE FROM ONGOING ANALYSIS, TREND INFO, OR OTHER ANALYSIS AS APPROPRIATE.

Description of Event or Problem · 1

DRUG/DILUENT: SOLUMEDROL WITH 5% GLUCOSE. FILL VOLUME: 250ML. FLOW RATE: 100ML/HR. PROCEDURE: IV THERAPY. CATHPLACE: PERIPHERAL IV. DATE OF SURGERY: (B)(6) 2014. THE CUSTOMER REPORTED A FAST FLOW WHILE USING AN I-FLOW HOMEPUMP 400MLX100ML, FILLED WITH 1G SOLUMEDROL IN 250ML'S OF 5% GLUCOSE SOLUTION THAT WAS CONNECTED TO A PT. THE SOLUTION WAS ADMINISTERED IN 1 HOUR & 20 MINUTES INSTEAD OF 2 HOURS & 30 MINUTES. THE INFUSION WAS VIA A PERIPHERAL IV. THE PUMP WAS PLACED IN THE FANNY PACK AROUND THE PT'S WAIST. THE INFUSION TEMPERATURE WAS ABOUT 24 DEGREES CELSIUS. INFUSION STARTED (B)(6) 2014 AT 9AM OR 10AM. INFUSION ENDED (B)(6) 2014 AT 10:30AM. INCIDENT NOTICED AT (B)(6) 2014 AT 10:30AM. THE PUMP WAS EMPTIED AT THE END OF THE INFUSION. SAMPLE WILL BE RETURNED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
203083 HOMEPUMP ECLIPSE: 400ML, 100ML ELASTOMERIC PUMP MEB I-FLOW, LLC E401000-10 0200974832

Patients

Seq Age Sex Outcome Treatment
1 49 YR