FDA Adverse Event Malfunction Summary report: N

DUAL COOLER/HEATER

MDR report key: 3853216 · Received April 3, 2014

Report

Report Number
1828100-2014-00237
Event Type
Malfunction
Date Received
April 3, 2014
Date of Event
March 13, 2014
Report Date
March 13, 2014
Manufacturer
TERUMO CARDIOVASCULAR SYSTEMS CORP.
Product Code
DWC
PMA / PMN Number
PREAMEND
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
VA, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

THE FIELD SERVICE REP (FSR) CONFIRMED THIS COMPLAINT. THE FSR FOUND THAT MOTOR SPEED RESISTOR "R2" WAS OPEN. HE REPLACED AND ADJUSTED THE NEW R2 FOR SPECIFICATION OF 35 OHMS. THE UNIT OPERATED TO MFR SPECIFICATIONS AND WAS RETURNED TO CLINICAL USE. THE SUSPECT RESISTOR R2 WAS RETURNED TO THE MFR FOR FURTHER EVALUATION. IF ADDITIONAL INFO BECOMES AVAILABLE ON THIS COMPLAINT THAT WOULD ALTER THE FACTS AND/OR CONCLUSION, A SUPPLEMENTAL REPORT WILL BE FILED ACCORDINGLY.

Description of Event or Problem · 1

THE FIELD SERVICE REP (FSR) REPORTED THAT WHEN PERFORMING REPAIR ON UNIT FOR MDR #1828100-2014-00221, HE FOUND THAT THE NEW PUMP WOULD NOT RUN IN COOLING MODE #2. SINCE THE EVENT OCCURRED DURING PREVENTATIVE MAINTENANCE, THERE WAS NO PT INVOLVEMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
203202 DUAL COOLER/HEATER SARNS DUAL COOLER/HEATER DWC TERUMO CARDIOVASCULAR SYSTEMS CORP. 11160

Patients

Seq Age Sex Outcome Treatment
1