FDA Adverse Event Malfunction Summary report: N

PROXIMATE CURVED STAPLER

MDR report key: 3853179 · Received June 6, 2014

Report

Report Number
3005075853-2014-03822
Event Type
Malfunction
Date Received
June 6, 2014
Date of Event
May 9, 2014
Report Date
May 13, 2014
Manufacturer
ETHICON ENDO-SURGERY, LLC.
Product Code
GDW
PMA / PMN Number
K983536
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NC, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE ANALYSIS RESULTS FOUND THAT THE DEVICE ARRIVED IN GOOD VISUAL CONDITION. ONLY ANVIL HALF WASHER WAS PRESENT AND CUT AND THERE WERE NO STAPLES PRESENT, INDICATING THAT THE DEVICE ACHIEVED A FULL FIRING STROKE. THE DEVICE WAS RELOADED WITH STAPLES, A NEW WASHER WAS PLACED ON THE DEVICE AND IT WAS TESTED FOR FUNCTIONALITY. IT FIRED AND FORMED ALL THE STAPLES AS WELL AS COMPLETELY CUT THE TEST MEDIA AND THE BREAKAWAY WASHER WITHOUT INCIDENT. THE STAPLE LINE WAS COMPLETE AND THE STAPLES WERE NOTED TO HAVE THE PROPER B-FORMED SHAPE. IT SHOULD BE NOTED THAT WHEN REMOVING THE DEVICE; OPEN THE INSTRUMENT BY TURNING THE ADJUSTING KNOB COUNTERCLOCKWISE. FOR EASY REMOVAL, ONLY OPEN THE INSTRUMENT ONE-HALF TO THREE-FOURTHS REVOLUTIONS. TO ASSURE THE ANVIL IS FREE FROM TISSUE, ROTATE THE INSTRUMENT 90° IN BOTH DIRECTIONS. TO WITHDRAW THE OPEN INSTRUMENT, GENTLY APPLY REARWARD TRACTION WHILE SIMULTANEOUSLY ROTATING. PLEASE REFERENCE INSTRUCTIONS FOR USE FOR ADDITIONAL INFORMATION NEEDED. THE BATCH RECORD WAS REVIEWED AND NO ANOMALIES WERE NOTED DURING THE MANUFACTURING PROCESS.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A LOW ANTERIOR RESECTION PROCEDURE, THE DEVICE WAS PLACED IN THE RECTUM AND FIRED. EVERYTHING WORKED PROPERLY. ONCE THE SURGEON ATTEMPTED TO REMOVE THE DEVICE IT GOT STUCK ON TISSUE. THEY TRIED TURNING THE DEVICE A COMPLETE 180 DEGREE. THEY DID EVENTUALLY GET IT TO RELEASE FROM THE TISSUE. HOWEVER, WHEN THE SURGEON WENT BACK TO THE TOP, THE ANASTOMOSIS FELL APART. THEY WERE ABLE TO DO ANOTHER ANASTOMOSIS AND COMPLETED THE CASE. THE PATIENT IS CURRENTLY FINE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
332698 PROXIMATE CURVED STAPLER STAPLE, IMPLANTABLE GDW ETHICON ENDO-SURGERY, LLC. NA L4E91C

Patients

Seq Age Sex Outcome Treatment
1