PROXIMATE CURVED STAPLER
Report
- Report Number
- 3005075853-2014-03822
- Event Type
- Malfunction
- Date Received
- June 6, 2014
- Date of Event
- May 9, 2014
- Report Date
- May 13, 2014
- Manufacturer
- ETHICON ENDO-SURGERY, LLC.
- Product Code
- GDW
- PMA / PMN Number
- K983536
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NC, US
- Reporter Occupation
- OTHER
Narratives
(B)(4). THE ANALYSIS RESULTS FOUND THAT THE DEVICE ARRIVED IN GOOD VISUAL CONDITION. ONLY ANVIL HALF WASHER WAS PRESENT AND CUT AND THERE WERE NO STAPLES PRESENT, INDICATING THAT THE DEVICE ACHIEVED A FULL FIRING STROKE. THE DEVICE WAS RELOADED WITH STAPLES, A NEW WASHER WAS PLACED ON THE DEVICE AND IT WAS TESTED FOR FUNCTIONALITY. IT FIRED AND FORMED ALL THE STAPLES AS WELL AS COMPLETELY CUT THE TEST MEDIA AND THE BREAKAWAY WASHER WITHOUT INCIDENT. THE STAPLE LINE WAS COMPLETE AND THE STAPLES WERE NOTED TO HAVE THE PROPER B-FORMED SHAPE. IT SHOULD BE NOTED THAT WHEN REMOVING THE DEVICE; OPEN THE INSTRUMENT BY TURNING THE ADJUSTING KNOB COUNTERCLOCKWISE. FOR EASY REMOVAL, ONLY OPEN THE INSTRUMENT ONE-HALF TO THREE-FOURTHS REVOLUTIONS. TO ASSURE THE ANVIL IS FREE FROM TISSUE, ROTATE THE INSTRUMENT 90° IN BOTH DIRECTIONS. TO WITHDRAW THE OPEN INSTRUMENT, GENTLY APPLY REARWARD TRACTION WHILE SIMULTANEOUSLY ROTATING. PLEASE REFERENCE INSTRUCTIONS FOR USE FOR ADDITIONAL INFORMATION NEEDED. THE BATCH RECORD WAS REVIEWED AND NO ANOMALIES WERE NOTED DURING THE MANUFACTURING PROCESS.
IT WAS REPORTED THAT DURING A LOW ANTERIOR RESECTION PROCEDURE, THE DEVICE WAS PLACED IN THE RECTUM AND FIRED. EVERYTHING WORKED PROPERLY. ONCE THE SURGEON ATTEMPTED TO REMOVE THE DEVICE IT GOT STUCK ON TISSUE. THEY TRIED TURNING THE DEVICE A COMPLETE 180 DEGREE. THEY DID EVENTUALLY GET IT TO RELEASE FROM THE TISSUE. HOWEVER, WHEN THE SURGEON WENT BACK TO THE TOP, THE ANASTOMOSIS FELL APART. THEY WERE ABLE TO DO ANOTHER ANASTOMOSIS AND COMPLETED THE CASE. THE PATIENT IS CURRENTLY FINE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 332698 | PROXIMATE CURVED STAPLER | STAPLE, IMPLANTABLE | GDW | ETHICON ENDO-SURGERY, LLC. | NA | L4E91C |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |