INRATIO2
Report
- Report Number
- 2027969-2014-00289
- Event Type
- Malfunction
- Date Received
- April 2, 2014
- Date of Event
- March 5, 2014
- Report Date
- March 12, 2014
- Manufacturer
- ALERE SAN DIEGO, INC.
- Product Code
- GJS
- PMA / PMN Number
- K092987
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
IT IS INDICATED THAT PRODUCT IS NOT RETURNING FOR EVALUATION. THEREFORE, INVESTIGATION OF THE COMPLAINT TO DETERMINE ROOT CAUSE CANNOT BE COMPLETED. SINCE THE PRODUCT ASSOCIATED WITH THE COMPLAINT WAS NOT RETURNED, RETAIN TESTING ON INRATIO STRIPS WAS PERFORMED DURING AN IN-HOUSE INVESTIGATION. RETAIN STRIP TESTING RESULTS MET BOTH ACCURACY AND REPEATABILITY CRITERIA. THE PRODUCTS PERFORMED AS EXPECTED AND PRODUCT DEFICIENCIES WERE NOT OBSERVED. ALTHOUGH THE ROOT CAUSE ANALYSIS DID NOT INCLUDE RETURN TESTING, IMPROPER USE OF THE METER AND IMPROPER TECHNIQUES WERE IDENTIFIED IN THE COMPLAINT. THESE COULD NOT BE RULED OUT AS A CAUSE OF THE UNEXPECTED RESULTS. THE MANUFACTURING RECORDS FOR THE LOT WERE REVIEWED. THE NON-CONFORMANCE ASSOCIATED WITH THIS LOT WAS NOT RELEVANT TO THE INITIAL COMPLAINT AND DOES NOT AFFECT PRODUCT PERFORMANCE. NO FURTHER INVESTIGATION WILL BE PURSUED AT THIS TIME. BASED ON THE INFORMATION AVAILABLE, THERE IS NO INDICATION OF A PRODUCT DEFICIENCY. NO CORRECTIVE ACTION IS REQUIRED AT THIS TIME.
CALLER ALLEGED IMPRECISION WITH INRATIO METER. RESULTS AS FOLLOWS: (B)(6) 2014: INRATIO = 1.4, 3.1; (B)(6) 2014: INRATIO = 5.0, 4.7; 2 HOURS BETWEEN TESTS ON (B)(6) 2014; 5 MINS BETWEEN TESTS ON 03/12/2014. THERAPEUTIC RANGE: 2.5-3.5. METER WAS IN PATIENT'S LAP WHILE TESTING ON (B)(6) 2014. METER WAS IN DAUGHTER'S HAND DURING APPLICATION, THEN WAS PLACED ON A TABLE ON (B)(6) 2014. METER NOT IN CORRECT MODE WHEN FINGERSTICK PERFORMED; MILKING FINGER AFTER FINGERSTICK; SAMPLE NOT APPLIED IMMEDIATELY AFTER FINGERSTICK; TOUCHED FINGER TO SAMPLE WELL.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 200528 | INRATIO2 | PROTHROMBIN TIME TEST | GJS | ALERE SAN DIEGO, INC. | 100071 | 333790 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |