FDA Adverse Event Malfunction Summary report: N

FUSION PRE-LOADED WITH ACROBAT WIRE GUIDE

MDR report key: 3853136 · Received April 2, 2014

Report

Report Number
1037905-2014-00113
Event Type
Malfunction
Date Received
April 2, 2014
Date of Event
March 3, 2014
Report Date
March 4, 2014
Manufacturer
WILSON-COOK MEDICAL INC.
Product Code
KNS
PMA / PMN Number
K052051
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UK
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 1

INVESTIGATION EVALUATION: A PRODUCT EVALUATION WAS NOT PERFORMED IN RESPONSE TO THIS REPORT BECAUSE THE PRODUCT SAID TO BE INVOLVED WAS NOT PROVIDED TO COOK FOR EVALUATION. THE REPORT COULD NOT BE CONFIRMED. THE DEVICE HISTORY RECORD FOR THE LOT NUMBER SAID TO BE INVOLVED WAS REVIEWED. A DISCREPANCY OR ANOMALY WAS NOT OBSERVED WITH THE PRODUCT THAT WAS RELEASED FOR DISTRIBUTION. INVESTIGATION CONCLUSIONS: WE COULD NOT CONDUCT A COMPLETE INVESTIGATION BECAUSE THE PRODUCT SAID TO BE INVOLVED WAS NOT RETURNED FOR EVALUATION. A DEFINITIVE CAUSE FOR THE REPORTED OBSERVATION COULD NOT BE DETERMINED. IF ADDITIONAL PRESSURE IS APPLIED TO THE WIRE GUIDE AND/OR ACCESSORY DEVICE(S) WHILE MOVING THE WIRE GUIDE INSIDE THE ACCESSORY DEVICE(S), THIS COULD CONTRIBUTE TO WIRE GUIDE DAMAGE. THE INSTRUCTIONS FOR USE FOR THIS PRODUCT NOTES FOR BEST RESULTS, WIRE GUIDE SHOULD BE KEPT WET. PRIOR TO DISTRIBUTION, ALL COOK FUSION PRE-LOADED WITH ACROBAT WIRE GUIDE SPHINCTEROTOME ARE SUBJECTED TO A VISUAL INSPECTION AND FUNCTIONAL TEST TO ENSURE DEVICE INTEGRITY. A REVIEW OF THE DEVICE HISTORY RECORD CONFIRMED THAT THE LOT SAID TO BE INVOLVED MET ALL MANUFACTURING REQUIREMENTS PRIOR TO SHIPMENT. CORRECTIVE ACTION: CORRECTIVE ACTION IS NOT WARRANTED AT THIS TIME BASED ON THE QUALITY ENGINEERING RISK ASSESSMENT. QUALITY ASSURANCE WILL CONTINUE TO MONITOR FOR COMPLAINT TRENDS AND REASSESS THE RISK ASSESSMENT RESULTS AS POST MARKET FEEDBACK CONTINUES TO BECOME AVAILABLE.

Description of Event or Problem · 1

DURING AN ENDOSCOPY PROCEDURE, A COOK FUSION PRE-LOADED WITH ACROBAT WIRE GUIDE SPHINCTEROTOME WAS USED. WHEN STRIPPING THE SPHINCTEROTOME THE WIRE GUIDE STARTED TO UNWIND AND THEN BROKE. NO SECTION OF THE DEVICE DETACHED AND BECAME FREE INSIDE THE PATIENT. A SECTION OF THE DEVICE DID NOT REMAIN INSIDE THE PATIENT'S BODY. THE PATIENT DID NOT REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE. ACCORDING TO THE INITIAL REPORTER, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
200196 FUSION PRE-LOADED WITH ACROBAT WIRE GUIDE KNS, UNIT, ELECTROSURGICAL, ENDOSCOPIC (WITH OR WITHOUT ACCESSORIES) KNS WILSON-COOK MEDICAL INC. W3330944

Patients

Seq Age Sex Outcome Treatment
1 OLYMPUS TJF 260 ENDOSCOPE