FDA Adverse Event Other Summary report: N

TROPHON EPR

MDR report key: 3853109 · Received May 1, 2014

Report

Report Number
3853109
Event Type
Other
Date Received
May 1, 2014
Date of Event
April 24, 2014
Report Date
May 1, 2014
Manufacturer
NANOSONICS INC.
Product Code
OUJ
Report Source
User Facility report
Reporter Location
SC, US

Narratives

Description of Event or Problem · 1

THE HEAT FROM THE TROPHON IS LOOSENING THE EPOXY THAT HOLDS THE VAGINAL PROBE TOGETHER. THE VAGINAL PROBES ARE FALLING APART. WE HAVE HAD 6 VAGINAL PROBES TO FALL APART SINCE USING THE TROPHON. THEREFORE, WE HAVE HAD TO PURCHASE NEW PROBES. THE TROPHON UTILIZES A HYDROGEN BASED CHEMICAL STERILIZATION SOLUTION. WE BELIEVE THIS MAY BE CONTRIBUTING TO THE LOOSENING OF THE EPOXY OF THE VAGINAL PROBES AND THUS CAUSING THEM TO BREAK. OUR FACILITY IS IN THE PROCESS OF RETURNING ALL 12 TROPHONS BACK TO THE MANUFACTURER.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
263604 TROPHON EPR HIGH LEVEL DISINFECTION REPROCESSING INSTRUMENT FOR ULTRASONIC TRANSDUCERS OUJ NANOSONICS INC. * *

Patients

Seq Age Sex Outcome Treatment
1 A VAGINAL PROBE