FDA Adverse Event Malfunction Summary report: N

INFINITI VISION SYSTEM OZIL

MDR report key: 3853105 · Received April 2, 2014

Report

Report Number
2028159-2014-00500
Event Type
Malfunction
Date Received
April 2, 2014
Date of Event
February 1, 2014
Report Date
February 14, 2014
Manufacturer
ALCON - IRVINE TECHNOLOGY CTR
Product Code
HQC
PMA / PMN Number
K082845
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MD, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

THE COMPANY REPRESENTATIVE EXAMINED THE SYSTEM AND CONFIRMED THE REPORTED EVENT. THE FLUIDICS CONTROLLER PRINTED CIRCUIT BOARD (PCB) WAS REPLACED. UNRELATED TO THE REPORTED EVENT, THE DVD-ROM DRIVE WAS ALSO REPLACED. THE SYSTEM WAS THEN TESTED AND MET ALL PRODUCT SPECIFICATIONS. THERE WAS NO SAMPLE RETURNED FOR EVALUATION AND NO ADDITIONAL INFORMATION PROVIDED RELATED TO THIS EVENT. FOR THIS REASON, STEPS COULD NOT BE TAKED NOT REPLICATE OR CONFIRM THE REPORTED EVENT. A REVIEW OF COMPLAINTS FOR THE LAST 24 MONTHS DID NOT INDICATE ANY ADDITIONAL SIMILAR REPORTS FOR THIS SYSTEM. THE ROOT CAUSE CANNOT BE DETERMINED CONCLUSIVELY. (B)(4).

Description of Event or Problem · 1

A CUSTOMER REPORTED DURING A PROCEDURE THAT THE SYSTEM DISPLAYED A SYSTEM MESSAGE. THE SURGERY WAS COMPLETED USING AN ALTERNATE SYSTEM WITH NO HARM TO THE PATIENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
199820 INFINITI VISION SYSTEM OZIL PHACOFRAGMENTATION SYSTEM HQC ALCON - IRVINE TECHNOLOGY CTR INFINITI OZIL NA