SYMPHONY
Report
- Report Number
- 1000165971-2014-00317
- Event Type
- Injury
- Date Received
- June 6, 2014
- Date of Event
- May 7, 2014
- Report Date
- May 7, 2014
- Manufacturer
- SORIN CRM
- Product Code
- NVZ
- PMA / PMN Number
- PP950029
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
REPORTEDLY, PATIENT RETURNED FOR A ROUTINE FOLLOW-UP BUT THE DEVICE COULD NOT BE INTERROGATED BY 2 DIFFERENT PROGRAMMERS (MESSAGE "INTERROGATION IS STOPPED¿ POPPED UP). MAGNET WAS PLACED ON THE DEVICE AND THE RATE WAS SHOWN TO BE 80MIN-1 WITH GOOD CAPTURE TEST AS WELL. CLINIC IS CONCERNED WHY THE SYMPHONY FLIPPED TO ELECTIVE REPLACEMENT INDICATOR SO SOON WHEN GOOD BATTERY CELL IMPEDANCE WAS SEEN AT LAST FOLLOW-UP (0.69 KOHMS A YEAR AGO). IT WAS POSSIBLE TO PROGRAM THE DEVICE TO NOMINAL MODE VIA TELEMETRY HEAD BUTTON, BUT STILL COULD NOT INTERROGATE THE DEVICE (SAME MESSAGE KEPT POPPING UP). THE PHYSICIAN DECIDED TO CHANGE THE DEVICE THAT WILL BE RETURNED FOR ANALYSIS.
REPORTEDLY, PATIENT RETURNED FOR A ROUTINE FOLLOW-UP BUT THE DEVICE COULD NOT BE INTERROGATED BY 2 DIFFERENT PROGRAMMERS (MESSAGE "INTERROGATION IS STOPPED" POPPED UP). MAGNET WAS PLACED ON THE DEVICE AND THE RATE WAS SHOWN TO BE 80MIN-1 WITH GOOD CAPTURE TEST AS WELL. CLINIC IS CONCERNED WHY THE SYMPHONY FLIPPED TO ELECTIVE REPLACEMENT INDICATOR SO SOON WHEN GOOD BATTERY CELL IMPEDANCE WAS SEEN AT LAST FOLLOW-UP (0.69 KOHMS A YEAR AGO). IT WAS POSSIBLE TO PROGRAM THE DEVICE TO NOMINAL MODE VIA TELEMETRY HEAD BUTTON, BUT STILL COULD NOT INTERROGATE THE DEVICE (SAME MESSAGE KEPT POPPING UP). THE PHYSICIAN DECIDED TO CHANGE THE DEVICE THAT WILL BE RETURNED FOR ANALYSIS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 331212 | SYMPHONY | PULSE GENERATOR, PERMANENT, IMPLANTABLE | NVZ | SORIN CRM | SYMPHONY SR 2250 | M061114 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Required Intervention |