LAG SCREWDRIVER GAMMA3® 380X110MM
Report
- Report Number
- 0009610622-2014-00269
- Event Type
- Malfunction
- Date Received
- June 6, 2014
- Date of Event
- May 22, 2014
- Report Date
- May 22, 2014
- Manufacturer
- STRYKER TRAUMA KIEL
- Product Code
- LXH
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- PHYSICIAN
Narratives
EVALUATION SUMMARY: THE RETURNED DEVICE MATCHED THE REPORT AND THE REPORTED EVENT WAS CONFIRMED. REVIEW OF THE MANUFACTURING RECORDS REVEALED NO DISCREPANCIES. THE ITEM RETURNED WAS DOCUMENTED AS FAULTLESS PRIOR TO DISTRIBUTION. ONE OF THE FOUR PEGS IS COMPLETELY BROKEN OFF. MATERIAL DISPLACEMENT IS VISIBLE. THE REMAINING PEGS ARE SIGNIFICANTLY BENT AND DEFORMED. THE TIP OF THE INNER ROD IS FURTHERMORE DEFORMED AND DAMAGED. THE BROKEN OFF AND DAMAGED PEGS ARE A RESULT OF AN INADEQUATE USAGE. THE APPEARANCE OF THE DAMAGE INDICATES THAT THE LAG SCREWDRIVER HAD BEEN OPERATED UNDER HIGH FORCES IN CCW / UNTIGHTENING DIRECTION, WHILST THE PEGS WERE NOT ENTIRELY ENGAGED IN THE SLOTS OF THE LAG SCREW. THE HEAVILY ABRADED AND PLASTICALLY DEFORMED THREAD WINDINGS ARE MOST LIKELY THE RESULT OF AN OBLIQUE ADAPTING UNDER HIGH FORCE APPLICATION. IN CASE OF INTENDED USE SUCH DAMAGE WOULD NOT HAVE OCCURRED. A PRE-DAMAGING OF THE INSTRUMENT IN PRIOR SURGERIES COULD NOT BE EXCLUDED. BASED ON THE ABOVE FACTS THE ROOT CAUSE OF THE REPORTED EVENT IS NOT RELATED TO A DEFICIENCY OF THE DEVICE, BUT IS RATHER LINKED TO IMPROPER HANDLING BY THE USER. NO NON-CONFORMITY WAS IDENTIFIED.
DEVICE WILL NOT BE RETURNED. IF ADDITIONAL INFORMATION BECOMES AVAILABLE IT WILL BE PROVIDED ON A SUPPLEMENTAL REPORT.
IT WAS REPORTED THAT THERE WAS A GAMMA NAIL REMOVAL AND CONVERSION TO A RIGHT TOTAL HIP DUE A NON-UNION. UPON ATTEMPTING TO REMOVE THE GAMMA NAIL THE LAG SCREW INSERTER GEAR WAS SHEARED OFF.
IT WAS REPORTED THAT THERE WAS A GAMMA NAIL REMOVAL AND CONVERSION TO A RIGHT TOTAL HIP DUE A NON-UNION. UPON ATTEMPTING TO REMOVE THE GAMMA NAIL THE LAG SCREW INSERTER GEAR WAS SHEARED OFF.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 332555 | LAG SCREWDRIVER GAMMA3® 380X110MM | INSTRUMENT | LXH | STRYKER TRAUMA KIEL | K0F30F9 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 61 YR |