ANIMAS VIBE
Report
- Report Number
- 2531779-2014-15989
- Event Type
- Malfunction
- Date Received
- June 5, 2014
- Report Date
- May 20, 2014
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- PATIENT
Narratives
FOLLOW-UP #1: DATE OF SUBMISSION 04/21/2015.DEVICE EVALUATION: THE DEVICE HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 04/09/2015 WITH THE FOLLOWING FINDINGS: ON EXAMINATION THE KEYPAD WAS INTACT WITHOUT DAMAGE. ON INVESTIGATION, THE OK BUTTON DEMONSTRATED INTERMITTENT UNRESPONSIVENESS. THE REMAINDER OF THE KEYPAD BUTTONS HAD NORMAL RESPONSE. THE KEYPAD COVER WAS REMOVED FOR INVESTIGATION AND REVEALED CONTAMINATION ON THE CONTACTS OF THE KEYPAD BUTTONS. INVESTIGATION DUPLICATED THE COMPLAINT. UNRELATED TO THE ORIGINAL COMPLAINT, THE DISPLAY WAS FOUND TO BE DIM AND DISCOLORED.
THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME. (B)(6).
ON (B)(6) 2014, THE REPORTER CONTACTED ANIMAS, ALLEGING A BUTTON/KEYPAD (TACTILE CHANGES/UNRESPONSIVE) ISSUE. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 328880 | ANIMAS VIBE | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |