ONETOUCHPING GLUCOSEMGMTSYSTEM
Report
- Report Number
- 2531779-2014-15978
- Event Type
- Malfunction
- Date Received
- June 5, 2014
- Report Date
- June 1, 2014
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- PATIENT
Narratives
THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.
FOLLOW-UP # 1 DATE OF SUBMISSION 06/30/2014 - DEVICE EVALUATION:THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 06/25/2014 WITH THE FOLLOWING FINDINGS: VISUAL INSPECTION CONFIRMED THAT THERE WAS MOISTURE BEHIND THE DISPLAY LENS. DURING TESTING, THE PUMP POWERED ON WITH AUDIO TONES AND VIBRATIONS USING THE RETURNED BATTERY CAP. THE KEYPAD COVER WAS INTACT; NO DAMAGE WAS OBSERVED. THE KEYPAD COVER WAS REMOVED AND NO CONTAMINATION WAS FOUND UNDER ANY BUTTON CONTACTS. A LEAK TEST WAS PERFORMED AND A DISPLAY LENS LEAK WAS FOUND. THE PUMP CASE WAS REMOVED AND EVIDENCE OF MOISTURE CONTAMINATION WAS FOUND INSIDE THE PUMP. UNRELATED TO THE COMPLAINT, EVALUATION REVEALED THAT THE DISPLAY WAS BLANK. THE COMPLAINT OF THE UNDER RESPONSIVE KEYPAD BUTTONS WAS NOT ABLE TO BE ADEQUATELY INVESTIGATED DUE TO THE BLANK DISPLAY AND MOISTURE DAMAGE.
ON (B)(6) 2014, THE REPORTER CONTACTED ANIMAS, ALLEGING A BUTTON/KEYPAD (TACTILE CHANGES W/MOISTURE) ISSUE. THE REPORTER ALLEGED THAT ALL KEYPAD BUTTONS WERE UNDER RESPONSIVE. IT WAS NOTED THAT THERE WAS MOISTURE BEHIND THE DISPLAY. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 328810 | ONETOUCHPING GLUCOSEMGMTSYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 21 YR |