ONETOUCHPING GLUCOSEMGMTSYSTEM
Report
- Report Number
- 2531779-2014-15966
- Event Type
- Malfunction
- Date Received
- June 5, 2014
- Report Date
- June 3, 2014
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- PATIENT
Narratives
THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.
(B)(4) DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 07/17/2014 WITH THE FOLLOWING FINDINGS: A REVIEW OF THE BLACK BOX SHOWED A REBOOT ON 06/03/2014. A VISUAL INSPECTION SHOWED EVIDENCE OF MOISTURE CONTAMINATION ON THE UNDERSIDE OF THE BATTERY CAP. THE BATTERY COMPARTMENT WAS INTACT. THE PUMP INTERMITTENTLY POWERED ON WITH THE RETURNED BATTERY CAP. A TEST CAP WAS USED; THE TEST CAP WAS ABLE TO FULLY TIGHTEN ON TO THE PUMP WITH NO POWER LOSS. THE PUMP WAS EXERCISED WITH THE TEST BATTERY CAP WITH NO REBOOTS, POWER LOSS, OR CALL SERVICE ALARMS. THE PUMP PASSED A LEAK TEST. THE PUMP COVER WAS OPENED AND NO EVIDENCE OF INTERNAL MOISTURE WAS FOUND. UNRELATED TO THE COMPLAINT, THE DISPLAY SCREEN WAS DIM AND DISCOLORED. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE.
ON (B)(6) 2014, THE REPORTER CONTACTED ANIMAS, ALLEGING A POWER (NO POWER) ISSUE. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 329141 | ONETOUCHPING GLUCOSEMGMTSYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 16 YR |