ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 3852487 · Received June 5, 2014

Report

Report Number
2531779-2014-15958
Event Type
Malfunction
Date Received
June 5, 2014
Report Date
May 30, 2014
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CO, US
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.

Additional Manufacturer Narrative · 1

DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2014 WITH THE FOLLOWING FINDINGS: ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE. THE BLACK BOX COULD NOT BE DOWNLOADED FOR REVIEW. DURING A VISUAL INSPECTION OF THE PUMP, MOISTURE WAS OBSERVED BEHIND THE PUMP DISPLAY LENS. THE PUMP WAS POWERED ON WITH AUDIBLE TONES AND VIBRATION; HOWEVER, THE DISPLAY SCREEN WAS BLANK. A LEAK TEST WAS PERFORMED AND CONFIRMED A DISPLAY LENS LEAK. THE PUMP CASE WAS REMOVED AND FURTHER EVIDENCE OF MOISTURE INGRESS WAS FOUND ON INTERNAL COMPONENTS.

Description of Event or Problem · 1

ON (B)(6) 2014, THE REPORTER CONTACTED ANIMAS, ALLEGING A CASING/CONDITION (MOISTURE INGRESS) ISSUE. IT WAS REPORTED THAT MOISTURE WAS VISIBLE BEHIND THE PUMP DISPLAY LENS. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
329139 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION

Patients

Seq Age Sex Outcome Treatment
1 14 YR