ONETOUCHPING GLUCOSEMGMTSYSTEM
Report
- Report Number
- 2531779-2014-15946
- Event Type
- Malfunction
- Date Received
- June 5, 2014
- Report Date
- June 1, 2014
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AZ, US
- Reporter Occupation
- PATIENT
Narratives
FOLLOW-UP # 1 DATE OF SUBMISSION 08/08/2014 ¿ DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2014 WITH THE FOLLOWING FINDINGS: VISUAL INSPECTION CONFIRMED THAT THE BATTERY COMPARTMENT WAS CRACKED. CORROSION WAS OBSERVED INSIDE THE BATTERY COMPARTMENT. A LEAK TEST WAS PERFORMED AND THE LEAK WAS FOUND AT THE CRACK IN THE BATTERY COMPARTMENT. THE PUMP CASE WAS REMOVED AND MOISTURE DAMAGE WAS FOUND INSIDE THE PUMP.
THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.
ON (B)(6) 2014, THE REPORTER CONTACTED ANIMAS, ALLEGING A CASING/CONDITION (MOISTURE INGRESS) ISSUE. THE REPORTER ALLEGED THAT THERE WAS MOISTURE/CORROSION IN THE BATTERY COMPARTMENT. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 328793 | ONETOUCHPING GLUCOSEMGMTSYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 32 YR |