FDA Adverse Event Malfunction Summary report: N

UNICEL DXH 800 COULTER CELLULAR ANALYSIS SYSTEM

MDR report key: 3852385 · Received June 5, 2014

Report

Report Number
1061932-2014-01213
Event Type
Malfunction
Date Received
June 5, 2014
Date of Event
May 7, 2014
Report Date
May 7, 2014
Manufacturer
BECKMAN COULTER
Product Code
GKZ
PMA / PMN Number
K081930
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SC, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

A BECKMAN COULTER (BEC) FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED TO ASSESS THE INSTRUMENT. THE FSE FOUND A HOLE IN TUBING GOING TO VALVE VL242 IN MANIFOLD 201 WHICH LEAKED FLUID INTO THE AIR MIX AND TEMPERATURE CONTROL (AMTC) CHAMBERS. THE FSE REPLACED THE TUBING GOING TO VALVE VL242 AND THE DIFFERENTIAL CHAMBER. THE FSE VERIFIED THE INSTRUMENT AND THE INSTRUMENT WAS RETURNED TO NORMAL OPERATION. (B)(4).

Description of Event or Problem · 1

THE CUSTOMER REPORTED APPROXIMATELY 10 MLS OF CLENZ AND SAMPLE LEAKED IN THE UNICEL DXH 800 COULTER CELLULAR ANALYSIS SYSTEM. THE LEAK WAS NOT CONTAINED WITHIN THE INSTRUMENT. THE CUSTOMER WAS WEARING GLOVES AND A LAB COAT AT THE TIME OF THE INCIDENT. THERE WAS NO REPORTED EXPOSURE TO MUCOUS MEMBRANES OR OPEN WOUNDS. NO ERRONEOUS PATIENT RESULTS WERE GENERATED IN CONNECTION WITH THIS EVENT. THERE WAS NO DEATH, INJURY, OR AFFECT TO USER OR TO PATIENT TREATMENT ATTRIBUTED TO THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
328651 UNICEL DXH 800 COULTER CELLULAR ANALYSIS SYSTEM COUNTER, DIFFERENTIAL CELL GKZ BECKMAN COULTER NA NA

Patients

Seq Age Sex Outcome Treatment
1