SEE H-10
Report
- Report Number
- 2023826-2014-00436
- Event Type
- Malfunction
- Date Received
- June 5, 2014
- Date of Event
- May 5, 2014
- Report Date
- May 7, 2014
- Manufacturer
- STAAR SURGICAL COMPANY
- Product Code
- HQL
- PMA / PMN Number
- P880091
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OH, US
- Reporter Occupation
- NURSE
Narratives
SILICONE ULTRAVIOLET-ABSORBING POSTERIOR CHAMBER THREE PIECE FOLDABLE INTRAOCULAR LENS (ELASTIMIDE®). (B)(4). EVALUATION CODES RESULTS (OTHER) VISUAL INSPECTION OF THE RETURNED LENS SHOWED ONE HAPTIC WAS TORN OFF AND PIECES OF THE OPTIC AND THE OTHER HAPTIC WAS TORN OFF AND MISSING. THERE WAS A REDDISH RESIDUE ON THE DEVICE. AN INVESTIGATION OF THE ROOT CAUSES OF LENS TEARS, SIMILAR TO THAT STATED IN THIS CLAIM, WERE ADDRESSED IN A CAPA OPENED, WHICH WAS SUBSEQUENTLY CLOSED. THE CAPA ADDRESSED DELIVERY SYSTEM ISSUES INCLUDING ALL STAGES OF MANUFACTURING OF THE INJECTORS AND CARTRIDGES. PROCESSES WERE REVIEWED AND REVISED AS OPPORTUNITIES FOR IMPROVEMENT WERE REVEALED. THE CAPA ALSO ADDRESSED HANDLING ERRORS. ALL INJECTOR/CARTRIDGE DIRECTIONS FOR USE (DFU) HAVE BEEN MODIFIED TO ADD FURTHER CLARIFICATIONS TO INSTRUCT THE USERS IN THE PROPER DELIVERY TECHNIQUES THAT ARE EFFECTIVE, AND MINIMIZE THE POTENTIAL FOR DAMAGING THE LENS. BASED ON THE COMPLAINT HISTORY AND THE EVALUATION OF THE RETURNED PRODUCT, POSSIBLE ROOT CAUSES FOR LENS TEARS INCLUDE BOTH DELIVERY SYSTEM ISSUES AND HANDLING ERRORS BY THE CUSTOMER. (B)(4).
THE CUSTOMER REPORTED THE AQ2010V SILICONE THREE PIECE LENS TORE UPON INSERTION. THE LENS WAS REMOVED AND ANOTHER SAME MODEL LENS WAS IMPLANTED WITHOUT ANY INJURY TO THE PATIENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 329047 | SEE H-10 | INTRAOCULAR LENS | HQL | STAAR SURGICAL COMPANY | AQ2010V | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 62 YR | AQ CARTRIDGE MODEL, LOT NUMBER 1293588| MSI-PM INJECTOR MODEL, LOT NUMBER UNKNOWN |