FDA Adverse Event Malfunction Summary report: N

SEE H-10

MDR report key: 3852307 · Received June 5, 2014

Report

Report Number
2023826-2014-00436
Event Type
Malfunction
Date Received
June 5, 2014
Date of Event
May 5, 2014
Report Date
May 7, 2014
Manufacturer
STAAR SURGICAL COMPANY
Product Code
HQL
PMA / PMN Number
P880091
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 1

SILICONE ULTRAVIOLET-ABSORBING POSTERIOR CHAMBER THREE PIECE FOLDABLE INTRAOCULAR LENS (ELASTIMIDE®). (B)(4). EVALUATION CODES RESULTS (OTHER) VISUAL INSPECTION OF THE RETURNED LENS SHOWED ONE HAPTIC WAS TORN OFF AND PIECES OF THE OPTIC AND THE OTHER HAPTIC WAS TORN OFF AND MISSING. THERE WAS A REDDISH RESIDUE ON THE DEVICE. AN INVESTIGATION OF THE ROOT CAUSES OF LENS TEARS, SIMILAR TO THAT STATED IN THIS CLAIM, WERE ADDRESSED IN A CAPA OPENED, WHICH WAS SUBSEQUENTLY CLOSED. THE CAPA ADDRESSED DELIVERY SYSTEM ISSUES INCLUDING ALL STAGES OF MANUFACTURING OF THE INJECTORS AND CARTRIDGES. PROCESSES WERE REVIEWED AND REVISED AS OPPORTUNITIES FOR IMPROVEMENT WERE REVEALED. THE CAPA ALSO ADDRESSED HANDLING ERRORS. ALL INJECTOR/CARTRIDGE DIRECTIONS FOR USE (DFU) HAVE BEEN MODIFIED TO ADD FURTHER CLARIFICATIONS TO INSTRUCT THE USERS IN THE PROPER DELIVERY TECHNIQUES THAT ARE EFFECTIVE, AND MINIMIZE THE POTENTIAL FOR DAMAGING THE LENS. BASED ON THE COMPLAINT HISTORY AND THE EVALUATION OF THE RETURNED PRODUCT, POSSIBLE ROOT CAUSES FOR LENS TEARS INCLUDE BOTH DELIVERY SYSTEM ISSUES AND HANDLING ERRORS BY THE CUSTOMER. (B)(4).

Description of Event or Problem · 1

THE CUSTOMER REPORTED THE AQ2010V SILICONE THREE PIECE LENS TORE UPON INSERTION. THE LENS WAS REMOVED AND ANOTHER SAME MODEL LENS WAS IMPLANTED WITHOUT ANY INJURY TO THE PATIENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
329047 SEE H-10 INTRAOCULAR LENS HQL STAAR SURGICAL COMPANY AQ2010V NA

Patients

Seq Age Sex Outcome Treatment
1 62 YR AQ CARTRIDGE MODEL, LOT NUMBER 1293588| MSI-PM INJECTOR MODEL, LOT NUMBER UNKNOWN