OXF UNI TIB TRAY SZA LM/RL PMA
Report
- Report Number
- 0001825034-2014-05244
- Event Type
- Injury
- Date Received
- June 5, 2014
- Date of Event
- May 14, 2014
- Report Date
- March 1, 2016
- Manufacturer
- BIOMET ORTHOPEDICS
- Product Code
- NRA
- PMA / PMN Number
- PP010014
- Removal / Correction Number
- N/A
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NC, US
- Reporter Occupation
- OTHER
Narratives
CURRENT INFORMATION IS INSUFFICIENT TO PERMIT A CONCLUSION AS TO THE CAUSE OF THE EVENT. REVIEW OF DEVICE HISTORY RECORDS SHOW THAT LOT RELEASED WITH NO RECORDED ANOMALY OR DEVIATION. THERE ARE WARNINGS IN THE PACKAGE INSERT THAT STATE THAT THIS TYPE OF EVENT CAN OCCUR: UNDER WARNINGS, NUMBER 1 STATES, "IMPROPER SELECTION, PLACEMENT, POSITIONING, ALIGNMENT AND FIXATION OF THE IMPLANT COMPONENTS MAY RESULT IN UNUSUAL STRESS CONDITIONS WHICH MAY LEAD TO SUBSEQUENT REDUCTION IN THE SERVICE LIFE OF THE PROSTHETIC COMPONENTS." UNDER POSSIBLE ADVERSE EFFECTS, NUMBER 10 STATES, "LOOSENING OR MIGRATION OF THE IMPLANTS CAN OCCUR DUE TO LOSS OF FIXATION, TRAUMA, MALALIGNMENT, BONE RESORPTION, EXCESSIVE ACTIVITY." THIS REPORT IS NUMBER 1 OF 2 MDRS FILED FOR THE SAME EVENT (REFERENCE 1825034-2014-05244 / 05245).
THIS FOLLOW-UP REPORT IS BEING FILED TO CORRECTED INFORMATION AND TO RELAY ADDITIONAL INFORMATION, WHICH WAS UNKNOWN AT THE TIME OF THE INITIAL MEDWATCH. THIS REPORT IS NUMBER 1 OF 2 MDRS FILED FOR THE SAME EVENT (REFERENCE 1825034-2014-05244 / 05245). NOT RETURNED BY PATIENT.
IT WAS REPORTED PATIENT UNDERWENT A LEFT PARTIAL KNEE ARTHROPLASTY ON (B)(6) 2013. SUBSEQUENTLY, PATIENT WAS REVISED ON (B)(6) 2014 DUE TO MALPOSITIONING OF FEMORAL AND TIBIAL COMPONENTS AND TIBIAL LOOSENING. ALL COMPONENTS WERE REMOVED AND A TOTAL KNEE SYSTEM WAS IMPLANTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 329509 | OXF UNI TIB TRAY SZA LM/RL PMA | PROSTHESIS, KNEE | NRA | BIOMET ORTHOPEDICS | N/A | 2957111 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Hospitalization| R |